Development of Unfolding Energy Spectrum with Clinical Linear Accelerator based on Transmission Data
facebook(new window)  Pirnt(new window) E-mail(new window) Excel Download
 Title & Authors
Development of Unfolding Energy Spectrum with Clinical Linear Accelerator based on Transmission Data
Choi, Hyun Joon; Park, Hyo Jun; Yoo, Do Hyeon; Kim, Byoung-Chul; Yi, Chul-Young; Min, Chul Hee;
  PDF(new window)
 Abstract
Background: For the accurate dose assessment in radiation therapy, energy spectrum of the photon beam generated from the linac head is essential. The aim of this study is to develop the technique to accurately unfolding the energy spectrum with the transmission analysis method. Materials and Methods: Clinical linear accelerator and Monet Carlo method was employed to evaluate the transmission signals according to the thickness of the observer material, and then the response function of the ion chamber response was determined with the mono energy beam. Finally the energy spectrum was unfolded with HEPROW program. Elekta Synergy Flatform and Geant4 tool kits was used in this study. Results and Discussion: In the comparison between calculated and measured transmission signals using aluminum alloy as an attenuator, root mean squared error was 0.43%. In the comparison between unfolded spectrum using HEPROW program and calculated spectrum using Geant4, the difference of peak and mean energy were 0.066 and 0.03 MeV, respectively. However, for the accurate prediction of the energy spectrum, additional experiment with various type of material and improvement of the unfolding program is required. Conclusion: In this research, it is demonstrated that unfolding spectra technique could be used in megavoltage photon beam with aluminum alloy and HEPROW program.
 Keywords
Linac;Unfolding energy spectra;Transmission measurement;Monte Carlo simulation;
 Language
Korean
 Cited by
1.
An effective dose assessment technique with NORM added consumer products using skin-point source on computational human phantom, Applied Radiation and Isotopes, 2016, 118, 56  crossref(new windwow)
 References
1.
Dong KR, Park BS, Kim SS, Kweon DC, Goo EH, Chung WK. Target motion analysis of the respiratory gated guided radiotherapy in liver cancer patients using 4D-CT. Journal of Radiation Protection and Research. 2010;35(2):63-68.

2.
Lee J, Kang J, Kim J, Kim B, Jeong D. Analysis by synthesis for estimation of dose calculation with gMOCREN and GEANT4 in medical image. Journal of Radiation Protection and Research. 2012;37(3):146-148. crossref(new window)

3.
Nath R, Biggs PJ, Bova FJ, Ling CC, Purdy JA, van de Geijn J, Weinhous MS. AAPM code of practice for radiotherapy accelerators: Report of AAPM Radiation Therapy Task Group No. 45. Med. Phys. 1994;21(7):1093-1121. crossref(new window)

4.
Stathakis S, Esquivel C, Quino LV, Myers P, Calvo O, Mavroidis P, Gutierrez AN, Papanikolaou N. Accuracy of the small field dosimetry using the Acuros XB dose calculation algorithm within and beyond heterogeneous media for 6 MV photon beams. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology. 2012;1:78-87. crossref(new window)

5.
Huang PH, Kase KR, Bjarngard BE. Spectral characterization of 4 MV bremsstrahlung by attenuation analysis. Med. Phys. 1981; 8(3):368-374. crossref(new window)

6.
Ahuja SD, Steward PG, Roy TS, Slessinger ED. Estimated spectrum of a 4-MV therapeutic beam. Med. Phys. 1986;13(3):368-373. crossref(new window)

7.
Catala A, Francois P, Bonnet J, Scouarnec C. Reconstruction of 12 MV bremsstrahlung spectra from measured transmission data by direct resolution of the numeric system AF=T. Med. Phys. 1995;22(1):3-10. crossref(new window)

8.
Shimozato T, Tabushi K, Kitoh S, Shiota Y, Hirayama C, Suzuki S. Calculation of 10 MV x-ray spectra emitted by a medical linear accelerator using the BFGS quasi-Newton method. Phys. Med. Biol. 2006;52(2):515.

9.
Yi CY, Chun KJ, Kim IJ, Kim BC, Kim JH et al. Development of measurement standard of medical ionizing radiation. 1711007782. Korea Research Institute of Standards and Science, Daejeon, Republic of Korea. 2015;233-235.

10.
Hinson WH, Bourland JD. Spectral reconstruction of high energy photon beams for kernel based dose calculations. Med. phys. 2002;29(8): 1789-1796. crossref(new window)

11.
Matzke M. The HEPROW program system. NEA-1666/01. Nuclear Energy Agency, Boulogne-Billancourt, France. 2004;18-25.