Advanced SearchSearch Tips
A Study on the Measurement Uncertainty of Flowmeter Calibrator
facebook(new window)  Pirnt(new window) E-mail(new window) Excel Download
 Title & Authors
A Study on the Measurement Uncertainty of Flowmeter Calibrator
Im, Gi-Won;
  PDF(new window)
The standard uncertainty of flowrate measurement is obtained by combining that of independent variables. Gravimetric and volumetric method were applied to determine the flowrate and the standard uncertainties of flowrate measurement by both methods were evaluated in accordance with the procedure recommended by International Organization for Standardization. The combined standard uncertainties of determining the flowrate were estimated from the sensitivity coefficient and the standard uncertainty of independent variables. For practical application, the methods for evaluating and expressing uncertainty in flow measurement were discussed. It was found that the uncertainties of the weighing and time measurement in gravimetric method, the volume and time measurement in volumetric method have dominant influence on that of flowrate measurement. With the quantitative analysis of the sensitivity coefficient, the contribution of the each variable uncertainty to the combined standard uncertainty of flowrate measurement is shown clearly.
Flowmeter Calibrator;Uncertainty;Flow Measurement;Calibration;Gravimetric Method;Volumetric Method;Sensitivity Coefficient;National Standard System;
 Cited by
곡관후단의 직관거리와 센서위치에 따른 초음파유량계의 오차특성,이동근;조용;

대한기계학회논문집B, 2011. vol.35. 8, pp.861-868 crossref(new window)
2개의 기준유량계를 이용한 유량계 교정장치의 측정불확도 평가에 관한 연구,최종오;이완규;임기원;

대한기계학회논문집B, 2004. vol.28. 10, pp.1219-1230 crossref(new window)
곡관의 하류에 설치된 전자기유량계의 유량신호 특성에 관한 실험적 연구(난류 유동),임기원;

대한기계학회논문집B, 2002. vol.26. 11, pp.1613-1621 crossref(new window)
코리올리스 질량유량계의 유량측정에 영향을 미치는 인자에 관한 실험적 연구,임기원;이완규;

대한기계학회논문집B, 2003. vol.27. 12, pp.1699-1707 crossref(new window)
기름 유량표준장치의 개발 및 측정 불확도에 관한 연구,임기원;최종오;

대한기계학회논문집B, 2003. vol.27. 8, pp.1071-1080 crossref(new window)
소형 코리올리 질량 유량계의 개발,임기원;지정근;

대한기계학회논문집B, 2006. vol.30. 6, pp.497-504 crossref(new window)
Build-Up 기법을 이용한 경질유 표준장치의 측정범위 확장,임기원;최종오;

대한기계학회논문집B, 2006. vol.30. 12, pp.1139-1146 crossref(new window)
Ki Won Lim, 2000, 'A Study on the Measurement Uncertainty of Pipe Prover,' J. of KSME(B), Vol. 24, No. 10, pp. 1388-1398

Ki Won Lim, Jong Seung Paik, 1986, 1987, 'Development of a Mobile Flow Proving System, Report of KRISS,' LRIS-86-36-IR, KRIS-87-42-IR

ISO, 1993(E), 'Guide to The Expression of Uncertainty in Measurement(1st edition),'

ISO 4185, 1980, 'Measurement of Liquid Flow in Closed Conduit-Weighing Methods,'

ISO 8316, 1984, 'Measurement of Liquid Flow in Closed Conduits-Method by Collection of the Liquid in a Volumetric Tank,'

JFI standard 1001, 1975, '液體用流量計 器着試驗方法(in Japanese),' 日本計量工業協會

Nak Sam Chung, Jong Oh Choi, 1999, 'KRISS Guide to The Expression of Uncertainty in Measurement, Report of KRISS,' KRISS-99-070-SP

Paul E. Pontius, 1974, 'Mass and Mass values,' NBS Monograph No.133

マテリアル·デ-タベ-ス(Material Database) 編集委員會, 1989, 'マテリアル·デ-タベ-ス(Material Database),' 日刊工業新聞社

Ki Won Lim, 1994, 'Standard Calibration Procedure of Electromagnetic Flowmeter,' KASTO-94-10-408-107

Ki Won Lim, Byung Ro Yoon, 2000, 'Development of the Characterization Technique for High Precise Grout Flowmeter, Report of KRISS,' KRISS--2000-23