Electrolytic Conductance Measurement using Four-Electrode Cell and Potentiometric Circuit

포텐티오메트릭 4-전극 용기에 의한 전도도 측정

  • 천정균 (서강대학교 이공대학 화학과) ;
  • 백운기 (서강대학교 이공대학 화학과)
  • Published : 1976.04.30

Abstract

A direct-reading conductance measuring system using a potentiostatic circuit and a four-electrode conductance cell was devised. The difficulties with the traditional method of using the Wheatston bridge and a two-electrode cell due to the complicated nature of the electrochemical system, the double layer capacitance and the Faradaic impedance at the electrodes, etc., could be avoided in this method. The devised instrument proved to be convenient and suitable for precise measurements. The results of measured conductivities of KCl and HCl solutions are reported.

Keywords

References

  1. Electrochemirtry J. Koryta;J. Dvorak;V. Bohackova
  2. Electrochemical Kinetics K.J. Vetter
  3. Physical Chemistry, An Advanced Treatise, LXA, Electrochemistry E. Yeager;J. Kuta;H. Eyring(ed.)
  4. J. Chem. Educ. v.48 J. Braunstein;G.D. Robbins
  5. J. Chem. Eng'g Data v.20 K.S. Kim;W. Paik
  6. The Physical Chemistry of Electrolytic Solutions H.S. Harned;B.B. Owen
  7. J. Amer. Chem. Soc. v.63 B.B. Owen;F.H. Sweeton
  8. J. Sci. Inster.(J. Phys. E) v.2 F.P. Anderson;H.C. Brookes;M.D. B Hotz;A.H. Spong
  9. Treatise on Eletrochemistry G. Kortum
  10. The Principles of Electrochemistry D.A. McInnes
  11. J. Chem. Phys. v.10 H.E. Gunning;A.R. Gordon
  12. J. Chem. Educ. v.50 T. Cyr;J. Prudhomme;M. Zador