Determination of Gallium by differential Pulse Adsorptive Stripping Voltammetry

펄스차이 흡착법김전압전류법에 의한 갈륨 정량

  • Published : 1992.12.20

Abstract

A very sensitive adsorptive stripping voltammetric method is studied on the gallium-morin complex at a hanging mercury drop electrode (HMDE) in 0.1 M acetate buffer solution. The effects of various analytical conditions are discussed on the reduction peak current of the adsorbed complex on the suface of HMDE. Interferences by other trace metals and surfactant are also discussed. Detection limit is 1.7 nM of gallium with 60 seconds deposition time, and the relative standard deviation (n = 7) at 4 ${\mu}$g/l gallium is 2.8%.

Keywords

References

  1. Anal. Sci. v.4 J. Ueda;C. Mizui
  2. Talanta v.37 F. Capitan;E. Manzano;J. L. Vilchez;L. F. Capitan-Valle
  3. Anal. Chim. Acta. v.185 J. Wang;J. M. Zadeii
  4. Anal. Chim. Acta. v.1 G. Charlot
  5. J. Korean Chem. Soc. v.35 S. C. Sohn;T. Y. Eom;Y. K. Ha;K.-S. Jung
  6. J. Electroanal. Chem. v.95 K. Ogura;Y. Enaka
  7. Anal. Chem. v.56 J. C. Yu;C. M. Wei
  8. Anal. Chem. v.47 E. D. Moorhead;P. H. Davis
  9. Anal. Chem. v.47 E. D. Moorhead;G. A. Forsberg
  10. Electroanalytical Chemistry v.16 J. Wang;A. J. Bard(Ed.)
  11. Anal. Chem. v.52 R. J. Nowak;F. A. Schultz;M. Umana;R. W. Murray
  12. Anal. Chem. v.36 M. H. Hulbert;I. Shain