Modification of Hydrogen Determinator for Total Hydrogen Analysis in Irradiated Zircaloy Cladding Tube

수소분석기 개조 및 조사후 지르칼로이 피복관의 총수소분석

  • Received : 1999.05.17
  • Published : 1999.12.25

Abstract

A hydrogen determinator was modified and installed in the glove box to analyse total hydrogen content in irradiated zircaloy tube. The analysis method of hydrogen is Inert Gas Fusion(IGF)-Thermal Conductivity Detection(TCD). The hydrogen recoveries of no tin method using Ti and Zr matrix standards, respectively, were available within $3{\mu}g$ of hydrogen. Also the smaller size of sample showed the better hydrogen recovery. It was found that the hydrogen standard of Ti matrix is avaliable to hydrogen analysis in zircaloy sample. The mean radioactivity of irradiated zircaloy sample was 10 mR/hr and hydrogen concentration was 130 ppm.

Keywords

hydrogen;glovebox;Inert Gas Fusion(lGF)-Thermal Conductivity Detection(TCD);irradiated zircaloy tube

References

  1. 分析化學 v.34 no.11
  2. Proceedings of seventh national symposium on mass spectrometry Y. S, Sayi;Prasad, Rajendra;K. I, Ramakumar;C. S, Yadav;P. S. Sankaran;G. C, Chappru;V, Aggarwal;S. K, Venugopal;H. C Jain;D. D, Sood
  3. Plata(Argentina) Scientific meeting of the Argentine Association of Nuclear Technology J. R. Granada;J. R, Santisteban
  4. Nuclear and Radiochemistry Symposium Y. S, Sayi;K. I, Prasad;Rajendra, Ramakumar;C. S, Yadav;P. S, Sankaran;G. C, Chappru;V, Venugopal;S. K, Agarwal;H. C, Jain;D. D, Sood
  5. Instruction manual for $LECO^R$ RH-404 Hydrogen Determinator $LECO^R$ Corporation
  6. Nuclear Instrument and Methods in Physics Research v.B 140 A. Bickel;C. Fred A. Sopchyshyn;Glenn H. McRae;Zane W. Walker;Lawrence Green
  7. Z. Anal. Chem. v.291 J. G. van Raaphorst;A. Kout;Fresenius
  8. Proceedings of a technical committee meeting on Recent developments in post-irradiation examination techniques for water reactor fuel T. Bredel
  9. 分析機器 v.14 no.9 和田行男;秋山
  10. 한국산업규격 KSD 2585-1995

Acknowledgement

Supported by : 과학기술부