Stationary and Moving Computed Radiography Grids : Comparative Observer's Perception

Computed Radiography에서 고정형 그리드와 이동형 그리드 영상의 인식률 비교

Lee, Kiho;Lee, Changhoon;Jin, Gyehwan
이기호;이창훈;진계환

  • Received : 2015.12.07
  • Accepted : 2015.12.25
  • Published : 2015.12.31

Abstract

This study assessed the degradation of image quality caused by grid artifacts and $moir{\acute{e}}$ pattern artifacts in a stationary grid, and the degradation of image quality caused by cut off artifacts in a moving grid. X-ray images were acquired in a stationary grid and a moving grid with X-ray exposure conditions of 100 cm, 80 kVp, and 30 mA using a CDRAD phantom and a 24 cm thickness acrylic phantom. Observer's perception of X-ray imaging using CDRAD Analyzer was mean 49.36, standard deviation 3.76, maximum 55.56, and minimum 38.67 in the stationary grid, and 47.04, 12.69, 55.56, and 20.89, respectively, in the moving grid. The stationary grid was superior to the moving grid in terms of the mean and standard deviation of observer's perception.

Keywords

Stationary Grid;Moving Grid;Perception;Computed Radiography;CDRAD Phantom;CDRAD Analyzer

References

  1. D.S. Kim, "Grid Angle Optimization and Grid Artifact Reduction in Digital Radiography Images Based on the Modulation Model", Journal of the Institute of Electronics and Information Engineers, Vol. 48, No. 3, pp.30-41, 2011.
  2. S.J. Lee, H.S. Cho, S.I. Choi, H.M. Cho, J.E.Oh, S.Y. Lee, , Y.O. Park and M.S. Lee, "Study on a moiré Artifact in the Use of Carbon Interspaced Antiscatter Grids for Digital Radiography", Journal of the Korean Society of Radiology, Vol. 2, No. 4, pp. 5-9, 2008.
  3. H.K. Hyun, S.H. Park, K.Y. Kim, H.M. Cho, and H. S. Cho, "Evaluation of Contrast-detail Characteristics of an A-Se Based Digital X-ray Imaging System", Journal of the Korean Society of Radiology, Vol. 1, No. 1, pp. 11-19, 2008.
  4. H.K. Hyun, S.H. Park, K.Y. Kim, H.M. Cho, and H. S. Cho, "Evaluation of Contrast-detail Characteristics of an A-Se Based Digital X-ray Imaging System", Journal of the Korean Society of Radiology, Vol. 1, No. 1, pp. 11-19, 2007.
  5. J.Y. Jung, H.S. Park, H.M. Cho, C.L. Lee, S.R. Nam, Y.J. Lee and H.J.Kim, "Imaging Characteristics of Computed Radiography Systems", Korean journal of medical physics, Vol. 19, No. 1, pp. 63-72, 2008.
  6. C.Y. Lin, W.J. Lee, S.J. Chen, C.H. Tsai, J.H. Lee, C.H. Chang and Y.T. Ching, "A Study of Grid Artifacts Formation and Elimination in Computed Radiographic Images", Journal of Digital Imaging, Vol. 19, No. 4, pp. 351-361, 2006. https://doi.org/10.1007/s10278-006-0630-8
  7. J.P. Hogge, C.H. Palmer, C.C. Muller, S.T. Little, D.C. Smith, P.P. Fatouros and E.S. Paredes, Quality assurance in mammography: artifact analysis, Radiographics, Vol. 19, No. 2, pp. 503-522, 1999. https://doi.org/10.1148/radiographics.19.2.g99mr13503
  8. E. Norrman, M. Gardestig, J. Persliden, H. Geijer, A clinical evaluation of the image quality computer program, CoCIQ, Journal of Digital Imaging, Vol. 18, No. 2, pp. 138-144, 2005. https://doi.org/10.1007/s10278-004-1036-0
  9. S. Rivetti, N. Lanconelli, M. Bertolini, D. Acchiappati, A new clinical unit for digital radiography based on a thick amorphous Selenium plate: Physical and psychophysical characterization, Medical Physics, Vol. 38, No. 8, pp. 4480-4488, 2011. https://doi.org/10.1118/1.3605471
  10. http://posterng.netkey.at/esr/viewing/index.php?module=viewing_pos ter&doi=10.1594/ecr2013/C-1619(2015.12.06.)

Acknowledgement

Supported by : National Research Foundation of Korea(NRF)