Analysis of the information capacity of digital radiograpy system

디지탈 X-선 촬영 시스템의 정보용량 분석

  • 김종효 (서울대학교 의과대학 의공학연구소) ;
  • 민병구 (서울대학교 의과대학 의공학교실) ;
  • 박광석 (서울대학교 의과대학 의공학교실) ;
  • 한만청 (서울대학교 의과대학 방사선학교실)
  • Published : 1990.10.01

Abstract

In this paper, the information capacity, the Integrate performance measure of digital radiography system has been quantitatively analyzed. The effect of various factors affecting the information capacity of the digital radiography system in multi-stage detection processes has been considered and analyzed In detail. The results show that the more Information capacity can be obtained with the smaller detector cell area, despite of the decreasing signal to noise ratio due to the reduced number of X-ray photons per detector cell. And also has been Indicated that the Increase of the information capacity may be severely lost by the addition of electric noise in final read-out stage.