Design of A Pendulum Type Motor-Driven Blood Pump for Artificial Heart

  • Chang, Jun-K. (Department of Biomedical Engineering, College of Medicine, Seoul National University) ;
  • Jung, Dae-Y. (Department of Biomedical Engineering, College of Medicine, Seoul National University) ;
  • Kim, Jong-W. (Department of Biomedical Engineering, College of Medicine, Seoul National University) ;
  • Min, Byoung-G. (Department of Biomedical Engineering, College of Medicine, Seoul National University) ;
  • Han, Dong-C. (Department of Mechanical Design & Production Engineering, College of Engineering, Seoul National University)
  • Published : 1990.05.23

Abstract

A new version of moving actuator electromechanical total artificial heart was designed to improve total efficiency, durability, and fitting inside thoracic cavity. As compared with our present type of the rolling cylinder actuator, this new model has a pendulum-type actuator with reciprocating motion around the fixed circular path, connected through the gear mechanisms to the motor. By using this mechanism, the efficiency and durability could be improved by replacing sliding mechanism with rolling contact elements. Also, the height of the pump could be decreased from 9cm to 7cm with static stroke volume 65cc. With these improvements, we have implanted this new pump in human size animal (less than 70Kg weight).