Abstract
A multi-parameter neuro muscular electrical simulator (NMES) system was developed to be used to find the optimal parameter condition in obtaining maximum muscle power and minimal fatigue. Since the performance of NMES is mainly determined by the characteristic of its output-stage circuit, we implemented 3 different circuits and compared output characteristics of them. Three amplifier circuits are; 1) a resonant switching converter, 2) a linear amplifier with a transformer, and 3) a step-up DC/DC converter with a high-voltage linear amplifier. Experimental results showed that the step up DC/DC converter with a high-voltage linear amplifier has the best performance.