Experimental Study of Second Harmonic Ultrasound imaging with a Weighted Chirp Signal

가중 쳐프 신호를 사용한 초음파 고조파 영상 기법의 실험적 고찰

  • Published : 2001.06.01

Abstract

In this Paper, a new harmonic imaging technique is proposed and evaluated experimentally. In the proposed method, a weighted chin signal with a hanning window is transmitted. The RF samples obtained on each array element are individually compressed by correlating with the reference signal defined as the 2nd harmonic (2f0) component of a transmitted chirp signal generated in a square-law system. The proposed method uses the compressed 2f0 component to form an image, for which the crosscorrelation term with f0 component should be suppressed below at least -60dB. After experiment, the 6dB pulse width and peak sidelobe level of the compressed 2f0 component were 0.7us and -60dB, respectively. This result shows that the proposed method can successfully eliminate the f0 component with a single transmit-receive event and therefore is more efficient than the conventional pulse inversion (PI) method in terms of frame rate. We also observed that the 2nd harmonic compont starts to decrease for source pressure higher than 210kPa in water, which implies that SNR of the 2nd harmonic imaging using short pulses cnanot be incresed beyond a certain limit.

Keywords