Determination of As(V) ion by Chemiluminescence Method

  • Lee, Sang-Hak (Department of Chemistry, Kyungpook National University) ;
  • Jeon, Hyun-Sook (Department of Sensor Engineering, Kyungpook National University)
  • Published : 2003.11.01

Abstract

A method to determine As(V) ions in aqueous solution by chemiluminescence method has been studied using a stopped flow system. The method is based on the increased chemiluminescence intensity with the addition of As(V) ion to a solution of lucigenin and hydrogen peroxide. The effects of KOH concentration, $H_2O_2$ concentration and flow rate of reagents on the chemiluminescence intensity have been investigated. The calibration curve for As(V) was linear over the range from $1.0{\times}l0^{-6}$M to $1.0{\times}l0^{-4}$M, the coefficient of correlation was 0.997 and the detection limit was $3.3{\times}l0^{-7}$M under the optimal experimental conditions.