초록

The objective of this study is to propose an experimental calibration facility in which a heat flux sensor can be calibrated under conductive condition by using helium gas. The heat flux calibration facility was designed, simulated and manufactured for use in a high heat transfer condition. It delivers heat fluxes up to a maximum of 35 KW $m^{-2}$. A copper block heated electrically with 3.5 KW power is designed to produce uniform temperature up to 600 K across its face. High heat fluxes are provided between hot plate and cold plate by 1 mm height helium filled gap. A cold plate is maintained around 300 K through pool boiling using a refrigerant and water-cooled heat exchanger. A simulation was conducted to verify the design of the main test section. To verify the performance of calibration facility, a heat flux sensor was examined. The measured heat fluxes were compared to the calculated one.

키워드