3-Dimensional Design of Gradient Coils for Magnetic Resonance Imaging

자기공명영상촬영용 경사자계코일의 3차원설계

  • 류연철 (고려대학교 대학원 전자정보학과) ;
  • 현정호 (고려대학교 대학원 의공학협동과정) ;
  • 이흥규 ((주)아이솔테크놀로지) ;
  • 오창현 (고려대학교 전자 및 정보공학부)
  • Published : 2006.10.27

Abstract

In this paper, the previous 2-D gradient coil design method using loop current elements is extended to 3-D or multi-layer structures which is useful for various MRI applications including MR microscopic imaging where relatively large space may be available for the implementation of the gradient coils. Either the power consumption or the stored energy (thus, inductance), or the combination of the two can be minimized with a set of chosen target field constraints. Complete 3-D design equations for the optimization as well as inductance or resistance calculation are derived. An effective coil shape correction method for a curved current pattern is also developed. The design method can also be easily extended to the active shielding structure.

Keywords