A Study on a Minimizing Method of Baseline Wandering in ECG

심전도 기저선 변동의 최소화방법에 관한 연구

  • 주장규 (관동대학교 전자통신공학과) ;
  • 김민규 (명지대학교 대학원 전자공학과) ;
  • 이기영 (관동대학교 정보통신공학부) ;
  • 김정국 (명지대학교 전자공학과)
  • Published : 2006.04.29

Abstract

In this paper, we propose a method to minimize the baseline wandering that make hard to detect R wave in ECG. This method uses a different signal between ECG and ascending slope tracing waves to minimize the baseline wandering. When the slope of ECG signal maintains the value or falls, the ascending slope tracing wave fellows ECG signal directly, and this wave holds that value of ECG signal when the slope begins to rises in a certain time(=hold time). After this hold time, this wave traces ECG signal again. To evaluate this minimizing method for baseline wandering, the experiments are carried out with 5 ECG data in the database of MIT/BIH. R waves in the proposed different signal are detected by using descending slope trace waves and compared with the annotation file. The results show that the proposed method Is sure to minimize the baseline wandering in ECG.