Implementation of Noninvasive Blood Pressure Measuring System for Home Health Care Using Oscillometric Method

오실로메트릭법을 적용한 홈헬스케어용 비침습적 혈압측정법의 구현

  • Published : 2006.05.01

Abstract

In this study, an implementation of a system for measuring more accurate blood pressure by non-invasive methods of oscillometric was performed to reduce errors and weaknesses of the existing invasive blood pressure measurement methods. The system is composed of pressure control, signal measurement and blood pressure signal processing units. To verify the validity of the system, tests of characteristics evaluations for pressure measurement unit, extraction of characteristic ratios for blood pressure estimation, blood pressure tracking by oscillometric method were performed. A group of five adult male was selected for the clinical test of the implemented system. The results of the oscillometric method in comparison with auscultatory method are that the maximum ratios of PAD of average, systolic and diastolic arterial pressure are 1.38%, 1.63% and 2.97% with SEP of 5.00, 3.72 and 4.34.

본 연구에서는 기존의 침습형 혈압 측정 방법의 위험성 및 불편성 등의 단점과 측정오차를 줄이고자 오실로메크릭법에 의한 혈압측정을 위하여 비관혈적인 방법으로 보다 정확하게 혈압을 추정할 수 있는 시스템을 구현하였다. 구현된 혈압 측정 시스템은 압력 제어부, 신호 측정부 그리고 혈압 신호처리부 등으로 구성하였다. 구현된 혈압측정 실험 시스템의 타당성을 검증하기 위하여 압력 측정부의 특성 평가 실험, 혈압 추정을 위한 특성비율 산출 실험, 오실로메트릭법에 의한 혈압 추정 및 평가 실험을 수행하였다. 본 연구에서 구현된 혈압 측정 실험 시스템으로 성인남성 5명을 실험 대상군으로 선정하여 데이터를 계측하고 혈압을 추정하는 실험을 한 결과 오실로메트릭에 의한 혈압 추정 및 평가 실험에서 청진법과의 평균동맥압, 수축기 혈압, 그리고 이완기 혈압에서 최대 절대편차 비율이 각각 1.38% 1.63 그리고 2.97% 이었고 최대 표준오차가 각각 5.00, 3.72, 그리고 4.34인 혈압 측정 시스템을 구현하였다.