Uncertainty and Compensation on the cell for Measurement of the Solid Permittivity Materials

고체 유전율 측정용 cell의 불확도 분석과 보상

  • Published : 2007.06.21

Abstract

For measurement of dielectric constants, the commercial parallel plate electrodes system with guard-ring electrode have been widely used up to now. The capacitance derived from the parallel plate electrodes capacitor with guard-ring electrode is calculated by the equation of ($C={\epsilon}\;{\cdot}\;\frac{area\;of\;electrod}{distance\;between\;electrodes}$). Therefore, in parallel plate electrode capacitor, the diameter of the guarded electrode, the gap size between guarded electrode and guard ring, and distance between two active electrode should be measured precisely to calculate dielectric constants from the measured capacitance. Consequently their mechanical measurement uncertainties are directly contributed. Especially the air-gap between the electrodes and dielectric specimen at the system must be existed and the measurement error derived from the air-gap is impossible to evaluate as measurement uncertainties. In this study, we analyze the uncertainty of the commercial dielectric constant test cell using 3 kinds CRMs.

Keywords