The Study for the analysis of the detection efficiency and the design of the radiation detector's collimator using MCNP

MCNP 기반 스테레오 방사선 검출기 콜리메이터 설계 및 선량검출효율 분석연구

  • Published : 2013.05.22

Abstract

The radiation sources leaked from large-scale radiation leak accident like the Hukushima nuclear power plant accident or nuclear explosions can cause to the very large damage for us. So that the damage can be minimized, we have being developed a detector that can providing information about the location of the source to remove dangerous substances quickly than the conventional single detector. In this paper, we designed the shielding and collimator for the development of the Stereo Radiation detector in order to detect contaminants using MCNP Code. And we analyzed the results that is detected from the discretionary position of the radiation source. The results of this paper will be used as the basis for designing efficient structure for the radiation detectors.

핵폭발이나 원자력 발전시설 사고와 같이 대규모 방사선사고 발생 시 주변지역은 감응방사선 또는 방사선 낙진으로 인해 오염된다. 이러한 오염지역을 원격에서 탐지하여 오염물질에 대한 분포 및 오염 정도를 확보한다면 오염물질 제거뿐만 아니라 오염에 대한 피해를 최소화 할 수 있다. 본 논문에서는 오염 물질을 탐지하기 위해 스테레오로 구성할 검출기 개발의 일환으로 MCNP코드를 이용하여 검출기의 차폐체 및 콜리메이터를 설계하고 임의 위치의 방사선원으로부터 검출되는 선량을 전산모사 후 결과를 분석하였다. 본 논문의 결과는 방사선 탐지를 위한 효율적인 검출기 구조를 설계를 위한 기초자료로 활용될 것이다.