Development of pulse diagnosis possible simulator using the stepper motor pumps

스텝 모터 펌프를 이용한 맥진 가능한 시뮬레이터의 개발

  • Published : 2016.10.27

Abstract

Virtual testing devices are required due to rapid changes in the health care industry and the increase of the medical or nursing workforce. The importance of devices such as the simulator, blood vessels, and lab equipment for modeling blood flow to the heart is increasing too. In this study, we made heart pump by using a step motor and developed device which simulates arterial, venous blood pressure, and blood flow. We finally evaluated the function of proposed device. The proposed system is composed of the pump for simulating, the valve device to describe the resistance of the artery and vein, and a reducing device showing the characteristics of the venous system. We used BOXER pump for heart simulator and silicon tube for arterial and venous vessels, and designed a reducing device. We also used the pressure sensor to measure arterial blood pressure. For the evaluation of the proposed system, we selected a range of 50~100mmHg of the blood circuit 60 per minute and then compared the blood pressure of a person and the measured blood pressure.

의료산업의 급속한 변화와 의료 및 간호 인력의 증가로 인하여 가상 시험 장치에 대한 개발이 요구되고 있으며 심장 모델링할 수 있는 시뮬레이터와 혈관, 혈류에 대한 실습장비에 대한 개발의 중요성이 증가하고 있다. 따라서 본 연구에서는 스텝 모터를 이용하여 심장 펌프를 제작하여 동맥과 정맥 혈압, 혈류 시뮬레이션을 위한 장비를 개발하고 그 기능을 평가하고자 한다. 제안 시스템은 심장 시뮬레이션을 위한 펌프와 동맥, 정맥 혈관의 저항을 모사하기 위한 밸브장치 그리고 정맥계의 특성을 나타내는 저감장치로 구성한다. 심장 시뮬레이터는 롱거펌프(BOXER)를 사용하였으며 동맥과 정맥혈관은 실리콘 튜브를 사용하고 저감장치를 설계 및 제작하였다. 그리고 동맥혈압 측정을 위하여 압력센서를 사용하였다. 또한 제안 시스템의 평가를 위하여 심장 박동수는 분당 60회로 혈압의 범위는 50 ~ 100mmHg로 선정하여 측정 혈압과 사람의 혈압을 비교 평가하였다.

Keywords