Journal of radiological science and technology (대한방사선기술학회지:방사선기술과학)
- Volume 3 Issue 1
- /
- Pages.81-86
- /
- 1980
- /
- 2288-3509(pISSN)
- /
- 2384-1168(eISSN)
A Study on the Comparision of Performance between Three-phase Full-wave and Single-phase Full-wave X-ray Generators
삼상(三相)12펄스 전파정류장치(全波整流裝置)와 단상(單相) 전파정류장치(全波整流裝置)의 성능(性能) 비교(比較)
- Kim, Young-Ill (Dept. of Radiotechnology, Shin Gu Junior College) ;
- Kim, Chang-Kyun (Dept. of Radiotechnology, Junior College of Public Health and Medical Technology, Korea University) ;
- Huh, Joon (Dept. of Radiotechnology, Junior College of Public Health and Medical Technology, Korea University)
- Published : 1980.07.18
Abstract
To compare three-phase 12-pulse full-wave X-ray generators with single-phase full-wave X-ray generators on their performance of outputs, authors studied the generating X-ray by means of exposure dose and radiographic density. The results were as follows; 1. The exposure doses of three-phase full-wave X-ray generators showed a 30%-60% increase as compared against of single-phase full-wave generators. 2. The transmitted doses of three-phase full-wave generators were more increased than single-phase full-wave X-ray generators. 3. To obtain the same density, 60kVp in three-phase full-wave generators were equivalent to
Keywords