A Study on the Construction of LDV System for a Measurement of the Fluid Velocity

유체속도 측정을 위한 레이저 도플러 유속계의 구성에 관한 연구

  • 최종원 (한국표준연구소 레이저연구실) ;
  • 조재흥 (한국표준연구소 레이저연구실) ;
  • 정명세 (한국표준연구소 레이저연구실)
  • Published : 1991.05.01

Abstract

The optics and the signal processor of dual beam laser Doppler velocimeter(LDV) was fabricated. By using the dual beam and the forward scattering, the optics part of LDV was fabricated. And the signal processor of LDV was designed by the frequency counter type using new 11:6 period timing device in order to remove error signals, and was made of the reference clock of a 500 MHz ECL oscillator. Doppler frequencies from 10KHz to 70MHz can be measured using the signal processor. In the accuracy of the period counting part, from 1.81x10**-4% to 1.27% is estimated, and in the accuracy of the validation logic part, from 0.78% to 14.78% is estimated.