Flow Dynamics Near End-To-End Anastomoses - Part I. In Vitro Compliance Measurement -

  • Kim, Y.H (Cardiovascular Fluid Dynamics Laboratory, School of Chemical Engineering, Georgia Institute of Technology)
  • Published : 1992.06.01

Abstract

Compliance mismatch across an end-to-end anastomosis was measured In the In vitro experimental setup. A 35mm camera was used and Image process was done in Gould/ DeAnza Image processor. The results showed that compliances of Penrose tubing and synthetic PTFE grafts were In good agreement with the previously reported In vivo data. PTFE grafts exhibited a nonlinear behavior with compliance decreasing with Increasing transmural pressure, whereas the compliance of the Penrose tubing remained relatively constant within the range of the pressures in which data were obtained. The lumen cross sections at the anastomosis were affected by the suture and the mismatch In compliance between the Penrose tubing and vascular grafts. The varla~lons In the lumen dtameter at the anastomosis was more pronounced with increasing transmural pressures. From the present study, it was clearly demonstrated that the compliance of prosthetic grafts Is much lower than that of the arteries. In addition to the hemodynamlc consequences, compliance mismatch across the anastomosis has been known to lead to Increased anastomotlc and suture stresses with resultant suture line dehlscence and false aneurysm formation. Thus, there are good hemodynamic reasons to suppose that Introduction of a less compliant arterial graft Into the arterial circulation wlll be damaging and that grafts should be made to match the elastic behavior of their host arteries as closely possible.

Keywords

References

  1. Graft materials in vascular surgery The role of mechanical properties in graft design Abbott,W.M.;Bouchier Hayes,D.J.;H.Dprdoch(ed.)
  2. J. Vasc. Surg. v.5 Effect of compliance mismatch on vascular graft patency Abbott,W.M.;Megerman,Hasson,J.E.;L'Italien,G.;Warnock,D.F.
  3. Surg. Forum v.27 Mismatch of mechanical properties as a cause of arterial prostheses thrombosis Clark,R.E.;Apostdou,S.;Kardos,J.L.
  4. Surg. v.72 Intimal and neointimal fibrous proliferation causing failure of arterial reconstruction Imparato,A.M.;Bracco,A.
  5. Arch. Surg. v.115 Matched elastic properties and successful arterial grafting Walden,R.;L'Italien,G.J.;Megerman,J.;Abbott,W.M.
  6. Acta. Chir. Scand. Suppl. v.529 Hemodynamic aspects of vascular grafting Nicolaides,A.N.
  7. Surg. Gyneol. and Obstetrics v.158 Effect of compliance alteration on healing of a poroud Dacron prosthesis in the thoracic aorta of the dog Cengiz,M.;Sauvage,L.R.;Berger,K.;Robel,S.B.;Robel,V.;Wu,H.;Walker,M.;Appleyard,R.F.;Wood,S.J.
  8. J. Vasc. Surg. v.2 Increased compliance near vascular anastomoses Hasson,J.,E.;Megerman,J.;Abbott,W.M.
  9. Arch. Surg. v.119 Postsurgical changes in arterial compliance Hasson,J.E.;Megerman,J.;Abbott,W.M.
  10. J. Vasc. Surg. v.3 Suture technique and perianastomotic compliance Hasson,J.E.;Megerman,J.;Abbott,W.M.
  11. J. Biomech. v.23 Predicting the compliance of small diameter vascular grafts from uniaxial tensile tests Stewart,S.F.C.;Lyman,D.J.
  12. Biomat. Med. Dev. Art. Org. v.11 Effect of healing on small internal diameter arterial graft compliance White,R.;Goldberg,L.;Hirose,F.;Shors,E.
  13. Ann. Surg. v.197 Downstream anastomotic hyperplasia : A mechanism of failure in Dacron arterial grafts LoGerfo,F.W.;Quist,W.C.;Nowak,M.D.
  14. J.Vasc. Surg. v.2 Effect of wall shear stress on initial thic-kening of arterially transplanted autoge-nous veins in dogs Morinaga,K.;Okadome,K.;Kuroki,M.;Miyazaki,T.;Muto,Y.;Inokuchi,K.
  15. Am. J. Physiol. v.233 Dynamic compliance of arterial grafts Baird,R.N.;Kidson,I.G.;L'Italien,G.J.;Abbott,W.M.
  16. Ann. Surg. v.175 Healing of arterial prosthesis in man : its incompleteness Berger,K.;Sauvage,L.R.;Rao,A.M.;Wood,S.J.
  17. Can. J. Surg. v.17 Necessity for elastic properties in synthetic arterial grafts Gozna,E.R.;Mason,W.F.;Marble,A.E.;Winter,D.A.;Dolan,F.G.
  18. J. Cardiovasc. Surg. v.21 Compliance : A continuing problem with vaxcular grafts Kinley,C.E.;Marble,A.E.