An experimental investigation of the stenosis effects on the pressure drop and flow change in the internal flow is presented. Stainless steel tubes of small diameter(3.175mm, 3.4mm) are used for the test section of the flow loop. Percent contraction ranges from 35% to 83% and the stenosis length ratio (L/d) is varied from 2.8 to 8. Water and aqueous glycerol solutions are used for Newtonian fluids and polymer solutions of Separan AP-273 (500 wppm, 1000 wppm) for non-Newtonian fluids. Pressure loss coefficients of non-Newtonian fluids decrease just as those of Newtonian fluids. The loss coefficients of Newtonian and non-Newtonian fluids increase as the percent contraction increases and the loss coefficients of non-Newtonian fluids are larger than those of Newtonian fluids for the same stenosed tube. The loss coefficient increases as the stenosis length ratio increases.