Fundamental Studies of the Electrical Characteristics of the Glow Discharge for the Development of HPLC Detector

글로우방전을 이용한 액체크로마토그라피 검출기 개발을 위한 기초특성연구

  • 이현주 (동덕여자대학교 약학과) ;
  • 김효진 (동덕여자대학교 약학과)
  • Published : 1995.08.01

Abstract

The plasma oscillation has been observed in an argon pressure between a tungsten anode and cathode consisting of an aqueous conducting solution. The effects of experimental parameters on the electrical characteristics of the glow discharge have been studied. The experimental parameters include the anode-cathode distance, pressure, methanol flow rate, and cathode materials. The glow discharge with liquid cathode and solid anode showed the potential sensitive detector for HPLC

Keywords

References

  1. Anal. Chem. v.63 Oscillating-plasma glow discharge as a detector for gas chromatography Kuzuya,M.;Piepmeier,E.H.
  2. Anal. Chem. v.66 Frequency, current and amplitude maps of oscillating-plasma glow discharge GC detector Cook,B.;Piepmeier,E.H.
  3. Anal. Chem. v.66 Fingerprint identification of organic compounds using an oscillating plasma glow discharge detector for gas chromatography Smith,D.L.;Piepmeier,E.H.
  4. Anal. Chem. v.67 Multivariate approach to fingerprint identification of organic compounds using an oscillating plasma glow discharge detector for gas chromatography Smith,D.L.;Piepmeier,E.H.
  5. Spectrochim Acta v.49B Studies of mechanisms in glow discharge oscillations Zhu,Z.;Piepmeier,E.H.
  6. Analytical Science & Technology v.6 no.4 Study on the plasma oscillation for gas chromatographic detector Kim,H.J.;Mang,D.Y.;Kang,J.S.
  7. Microchemical Journal Fundamental study and development of oscillating-plasma glow discharge GC detector Mang,D.Y.;You,S.Y.;Cha,B.C.;Kim,H.S.;Kim,H.J.
  8. Microchemical Journal Analytical characteristics of organic compounds using an oscillating plasma glow discharge detector for GC Mang,D.Y.;You,S.Y.;Cha,B.C.;Kim,H.S.;Lee,G.H.;Kim,H.J.