The Development and It′s Characteristics of New Film Dosimetry Algorithm for Personal Dosimetry

개인피폭 선량 측정을 위한 필름 배지 선량계의 새로운 알고리즘 개발 및 특성

  • 이병용 (울산대학교 의과대학, 서울중앙병원 치료방사선과학교실) ;
  • 장혜숙 (울산대학교 의과대학, 서울중앙병원 치료방사선과학교실) ;
  • 봉정균 (경기대학교 물리학과) ;
  • 권수일 (경기대학교 물리학과)
  • Published : 1995.12.01

Abstract

Purpose: We have developed new film dosimetry algorithm for personal dosimetry and examined its characteristics. Materials and methods: Agfagaevart personal monitoring 2/10 films are used. Films which are in the film badges filtered with Cu 0.3mm, plastic 1.5mm, Aluminum 0.6mm and tin 0.8mm, were exposed by standard dosimetry laboratory. Irradiated energy categories are ANSI N13.1l Category III, and IV. Manual type film precessor and X-rite film densitometor was used. Filtered densities to energy relations and does to transformed densities relations can be obtained ofter transformation of H&D curves to linear shape by polynomal fitting. Reults : Personal dose be determined within 25% error for category m and 15% for category IV. And we are able to evaluate the exposed energy. Conclusion : New algorithm developed in this study is good for personal dosimetry within 30% error range for catergory III and IV. It is expectd to be complete personal dosimetry algorithm with further study for categrory, I, Dand II V.

목적 : 필름배지를 이용하여 개인 피폭 선량을 측정하는 방법을 개발하고 그 특성을 알아보고자 한다. 재료 및 방법 : 필름으로는 Agfagaevart Personal monitoring 2/10 을 사용하였고, 필터로는 구리 0.3mm, 플라스틱 1.5mm, 알루미늄 0.6mm, 주석 0.8mm를 사용하였으며, 필름은 표준기관에서 교정하였다. 사용에너지는 ANSI N13.11 Category III, IV였고, 현상기는 수동현상기를, 농도계는 Xrite 농도계를 사용하였다. H&D 곡선을 선량에 따라 구한 뒤 다항식 전개를 이용하여 선량에 대해 직선성을 갖도록 변환하였다. 이후 필터와 방사선 에너지 관계를 구하여 선량 및 에너지를 추정할 수 있는 알고리즘을 개발하였다. 결과 : 본 연구에서 개발된 선량 평가 알고리즘은 해당 Category 의 전에너지에 대해 30% 이내의 정확도를 만족시켜 이 알고리즘이 개인 피폭 선량 측정에 이용될 수 있음을 알았다. Category I, II , V에 대하여 보완한다면 완벽한 선량 평가 알고리즘이 될 것으로 기대된다.

Keywords