Dosimetric Characteristics of Dynamic Wedge Technique
Dynamic Wedge의 조직내 방사선량 분포의 특성
- Oh Young Taek (Department of Radiation Oncology, Yonsei University College of Medicine) ;
- Keum Ki Chang (Department of Radiation Oncology, Yonsei University College of Medicine) ;
- Chu Seong Sil (Department of Radiation Oncology, Yonsei University College of Medicine) ;
- Kim Gwi Eon (Department of Radiation Oncology, Yonsei University College of Medicine)
- 오영택 (연세대학교 의과대학 치료방사선과학교실) ;
- 금기창 (연세대학교 의과대학 치료방사선과학교실) ;
- 추성실 (연세대학교 의과대학 치료방사선과학교실) ;
- 김귀언 (연세대학교 의과대학 치료방사선과학교실)
- Published : 1996.12.01
Abstract
Purpose : The wedge filter is the most commonly used beam modifying device during radiation therapy Recently dynamic wedge technique is available through the computer controlled asymmetric collimator, independent jaw. But dosimetric characteristics of dynamic wedge technique is not well known. Therefore we evaluate dosimetric characteristics of dynamic wedge compared to conventional fixed wedge. Materials and Methods : We evaluated dosimetric characteristics of dynamic wedge and fixed wedge by ion chamber, film dosimetry and TLD in phantoms such as water, polystyrene and average breast phantom. Six MV x-ray was used in
목적 : 방사선 치료에 있어서 조직내 등선량 분포곡선을 변형시킬 목적으로 쐐기 차폐물이 사용되고 있는데 최근 기존의 고정 쐐기와는 다르게 비대칭적인 콜리메이터인 Independent Jaw에 의해 등선량 분포곡선을 변형시키는 동적 쐐기 기법이 실용화 되고 있으나 아직 그 방사선 물리학적인 특성에 대해서는 잘 알려져 있지 않다. 이에 본 저자는 기존의 고정 쐐기와 비교하여 조직내 방사선량 분포의 특성을 알아보고자 본 연구를 계획하였다. 대상 및 방법 : 물 판톰, 폴리 스타이린 판톰, 평균 유방 모형 판톰을 대상으로 전리함, 필름, TLD 등을 이용하여 동적 쐐기와 고정 쐐기의 선량 분포를 측정하여 비교하였다. 방사선원은 선형 가속기의 6MV x선을 사용하였고