Measurements of Auditory Evoked Neuromagnetic Fields using Superconducting Quantum Interference Devices

SQUID를 이용한 뇌 청각유발 자장의 측정

  • 이용호 (한국표준과학연구원 초전도그룹) ;
  • 권혁찬 (한국표준과학연구원 초전도그룹) ;
  • 김진목 (한국표준과학연구원 초전도그룹) ;
  • 박용기 (한국표준과학연구원 초전도그룹)
  • Published : 1997.12.01

Abstract

Magnetic field sensors made from superconducting quantum interference device (SQUID) are the most sensitive low-frequency sensors available, enabling measurements of extremely weak magnetic fields from the brain. Neuromagnetic measurements allow superior spatial resolution, compared with the present electric measurements, and superior temporal resolution, compared with the fMRl and PET, providing useful informations for the functional diagnoses of the brain. We developed a 4-channel SQUID system for neuromagnetic applications. The main features of the system are its simple readout electronics and compact pickup coil structure. A magnetically shielded room has been constructed for the reduction of environmental magnetic noises. The developed SQUID system has noise level lower than the magnetic noise from the brain. Magnetic field signals of the spontaneous r-rhythm activity and auditory evoked magnetic fields have been measured.

초전도양자간섭소자(SQUID)를 이용한 자장센서는 현재 개발된 자장센서중에서 감도가 가장 우수한 소자로서 인체의 두뇌에서 발생하는 매우 미약한 자장에 측정이 가능하다. 뇌자도측정은 현재 많이 사용되고 있는 전기적인 측정(뇌파, 뇌유발전위) 에 비해 공간분해능이 우수하고, fMRI나 PET에 비해서는 시간분해능이 우수하므로 뇌기능연구에 유용하게 사용될 수 있다. 본 연구에서는 뇌자도 측정을 위하여 4-채널 SQUID시스템을 개발하였다. 개발된 시스템의 주요 특징은 새로운 방식의 SQUID센서를 채택함으로서 간단한 회로로써 SQUID구동이 가능하도록 하였으며, 검출코일의 신뢰성을 향상시키기 위하여 집적화된 평면형 코일을 사용하였다. 외부 환경잡음을 소거하기 위하여 자기차폐실을 설치하였고, 개발된 SQUID 시스템을 이용하여 뇌의 청각령으로부터 발생하는 자기신호를 측정하였다.

Keywords

References

  1. Advancess in biomagnetism Fundamentals of neuromagnetism G.L. Romani
  2. Rev. Mod. Phys. v.65 no.2 Magnetoencephalography-theory. instrumentation, and applications to noninvasive studies of the working human brain M. Hamalainen;R. hari;R. J. ilmoniemi;J. Knuutila;O.V. Lounasmaa
  3. J. Clinical Neourophysiology v.8 no.2 On brain's magnetic responses to sensort stimuli R. Hari
  4. IEEE Trans. Appl. Supercond v.5 no.2 SQUID magnetometers for biomagnetism and dnondestructive testing:Important questions and initial answers J.P. Wilkswo
  5. Science v.215 no.4533 Positron tomography and nuclear magnetic resonance imaging G.L. Brownell;T.F. Budinger;P.C. Lauterbur;P.L. McGeer
  6. Physica B v.197 SQUID technology and brain research O.V. Lounasmaa;J. Knuutila;R. Salmelin
  7. Proc. IEEE v.70 no.8 Principles and applications of SQUIDs J. Clarke
  8. Supercond. Sci. Tech. v.4 Dc SQUID systems overview D. Drun
  9. IEEE Trans. Appl. Supercond v.3 no.1 Relaxation oscillation SQUIDs with high δ V/δφ D.J. Adelerhof;H. Nijstad;J. Flokstra;H. Rogalla
  10. Supercond. Sci. Technol. v.9 An intergrated planar gradiometer based on a double relaxation oscillation SQUID Y.H.Lee;H.C.Kwon;J.M.Kim;Y.K.Park;J.C.Park
  11. J. Appl. Phys. v.58 no.11 Design of improved integrated thinfilm planar dc-SQUID gradiometers M.B.Ketchen
  12. IEEE Trans. Appl. Supercond v.7 no.2 A compact planar gradiometer systems for measuring tangential components of biomagnetic fields Y.H.Lee;H.C.Kwon;J.M.Kim;Y.K.Pakr;J.C.Park
  13. Rev. Sci. Instrum. v.66 no.10 Multichannel SQUID system detecting tangential components of cardiac magnetic field K.Tsukada (et al.)
  14. 한국자기학회지 v.6 no.4 생체자기계측을 위한 자기차폐실의 설치 및 특성 이용호;권혁찬;김진목;임청무;이상길;박용기;박종철