The Study on Measurement of Relative Conversion Factor in X-ray Image Intensifier

X선영상증배관의 상대변환계수 측정에 관한 검토

  • Kim, Sung-Chul (Institute of Radiological Technology, Dong A X- ray Co., LTD.) ;
  • Shin, Sung-Ill (Institute of Radiological Technology, Dong A X- ray Co., LTD.) ;
  • Lee, Sun-Sook (Institute of Radiological Technology, Dong A X- ray Co., LTD.) ;
  • Huh, Joon (Institute of Radiological Technology, Dong A X- ray Co., LTD.) ;
  • Kim, Sung-Soo (Dept. of Radiologic Technology, Shin Gu College)
  • 김성철 (동아엑스선기계 방사선기술연구소) ;
  • 신성일 (동아엑스선기계 방사선기술연구소) ;
  • 이선숙 (동아엑스선기계 방사선기술연구소) ;
  • 허준 (동아엑스선기계 방사선기술연구소) ;
  • 김성수 (신구전문대학 방사선과)
  • Published : 1997.12.20

Abstract

For the Evaluation of X-ray image intensifier, we measured radiation dose at input of I. I., brightness and fluorescence at output of I. I. by using X-ray exposure meter, optometer and fluorescence meter for the relative conversion factor. Especially, by using fluorescence meter, we could easily get relative conversion factor without having regulated machine by JIS. Since using, the quality of image intensifier is going down. Consequently, it needs continuous quality maintenance.