A Study on the Chest Radiography with Diseases in Consideration of Image Qualify and Patient Exposure

흉부질환의 화질과 피폭을 고려한 촬영조건의 연구

  • Lee, Man-Koo (Dept. of Radiotechnology, Wonkwang Health College) ;
  • Hayashi, Taro (The Chemotherapy Institute Kaken Hospital) ;
  • Ishida, Yuji (The Chemotherapy Institute Kaken Hospital)
  • 이만구 (원광보건전문대학 방사선과) ;
  • 임태랑 (일본화학요법연구소부속병원) ;
  • 석전유치 (일본화학요법연구소부속병원)
  • Published : 1997.12.20

Abstract

To evaluated the image quality and the patient exposure for the chest radiography, its fundamental imaging properties were investigated. The basic imaging properties were evaluated by measuring characteristic curves, relative speeds, average gradient, and patient exposure dose. The image qualities were evaluated by using a rotating meter. It was found that the film gradient of SRO750/SRH system was 3.13 and that of SRO750/HR-C30 was 1.77. The ratio of SRO1000/TMH to FS/RP-1 was 1 to 18.2. It was possible to visualize the static image when the exposure time was less than 2.5 msec in patient that had respiratory excessive motion, heart beat and natural physical motion, and less than 8.5 msec in normal. The ratio of medical exposure dose compared with our method was 1 to 25 in the routine chest radiography and maximum was 1 to 70. In estimation of the image quality in same cases, we found that the image of SRO1000/SRH and TMH of super sensitive systems was the same results. We found that these super sensitive screen-film systems were available for the chest radiography considering the relationship between the image quality and patient exposure.

Keywords