A Study of in-vitro Performances of the Intracardiac Axial Flow Pump

심장내 이식형 축류 혈액펌프의 in-vitro특성에 관한 연구

  • Published : 1998.02.01

Abstract

The intracardiac axial flow pump has been developed This device has several advantages: it fits well anatomically, its blood-contacting surface is small, and it is implanted as easily as an artificial heart valve replacement. The axial flow pump consists of an impeller and a motor, both of which are encased in a housing. Two types of impeller with 4 vanes and 6 vanes are used. Sealing of the motor shaft is achieved by means of a ferrofluidic seal. A flow of 5$\ell$/min was obtained at a differential pressure of 100mmHg with a motor speed of 7091rpm with the 4-vane impeller and 6402rpm with the 6-vane impeller. Sealing was kept against a pressure of 150mmHg at 7000rpm with the 4-vane impeller and 6402rpm with the 6-vane impeller. Sealing was kept against a pressure of 150mmHg at 7000rpm over 24 hours. The index of hemolysis was 0.056 with the 4-vane impeller and 0.214 with the 6-vane impeller. The intracardiac axial flow pump is a very promising circulatory support.

심장내 이식형 축류 혈액 펌프의 개발을 시도하였다. 이 펌프는 해부학적인 적합성이 좋고, 혈액과 접촉하는 면적이 적으며, 심장판막의 위치에 이식할 수 있어 이식술이 용이한 등의 이점이 있다. 축류혈액펌프는 입펠러와 소형 모터에 의해 구성되었으며, 모터축과 펌프의 본체의 밀봉은 자성유체 축봉이 이용되었다. 본 연구에서 개발한 혈액펌프의 펌프 특성은 4매 임펠러의 경우 회전수 7091[rmp], 6매 임펠러의 경우 회전수 6402[rpm]에서 유량 5[l/min], 차압 100[mmHm]을 얻을 수 있었으며, 자성유체 축봉의 내압 특성은 회전수 7000[rpm], 압력ㄹ 150[mmHg]에서 24시간 이상의 내구성을 얻을 수 있었다. 그리고, 용혈에 대하여 검토한 결과 4매 임펠러의 1H 는 0.0561[g/100L]이고, 6매 임펠러의 경우는 1H=0.214[g/100L]을 얻을 수 있었다. 이상의 실험결과는 심장내 이식 축류형 혈액펌프의 실현 가능성을 보여 주는 것으로 앞으로 큰 기대를 걸수 있다고 생각된다.

Keywords

References

  1. 日本機械技術硏究所資科 no.92 非拍動人工心臟の開發に關する調査硏究 山根志;筒井達夫
  2. Transactions of American Society for Artificial Internal Organs v.29 Physiopathological studies of nonpulsatile blood flow in chronic models I. Yada;L. R. Golding;H. Harasaki;G. Jacobs;S. Koike;R. Yozu;N. Sato;L. K. Fujimoto;J. Snow;E. Olsen;S. Murabayasi;V. S. Venkatesen;R. Kiraly;Y. Nose
  3. NTIS Report, NO1-HV-72975 Development of a reversing electrohydralic energy converter for the left ventricular assist device R. K. Jarvivik;M. S. Isaacson;A. P. Lioi;S. D. Nisen;J. Orth;K. Carr;T. R. Kessler;G. F. Wislicenus;D. B. Olsen;W. J. Kolff
  4. Centrifugal and axial flow pump A. J. Stepanoff
  5. 日本國特許 #S62-177369 液體用磁性流體軸 ツ-ル裝置 山丈政治;確井和法;黑岩顯彦
  6. 磁性流體 -基礎と應用- 武富荒;近角總信
  7. Artificial Organs v.18 no.1 The need for standardizing the index of hemolysis K. Naito;K. Mizuguchi;Y. Nose
  8. Proc. Internal Workshop Rotary Blood Pumps Hematological Principles of Hemolysis and Thrombosis with Special Reference to Rotary Blood Pumps Wurzunger;R.Opitz
  9. 日本 人工臟器 v.21 no.2 心臟弁位置理翊型 血液ポンプ(Valvo-Pump) の開發 山崎健;岡本英治;山本克之;三田村好矩;四津良平;田中隆
  10. Artificial Organs v.18 no.1 Development of the Baylor/NASA axial flow ventricular assist device:In vitro performance systematic hemolysis test results K. Mizuguchi;G. A. Damm;R. J. Bozean;J. W. Akkerman;G. S. Aber;P. AQ. Svejkovsky;J. W. Bacak;Y. Orime;S. Takatani;Y. Nose;G. P. Noon;M. E. Debakey