흉부 팬텀 CR 영상의 화질평가 및 피폭선량 경감에 관한 연구

  • Published : 1998.12.18

Abstract

With this experiment we have conclusion as follows. $\cdot$ Total FCR-9501 are 8 units. $\cdot$ All hospitals under study are conducting the high kvp radiographing with 115-120kvp and 4$\sim$10mAs. $\cdot$ The mean value of the chest exposure was 0.524$\sim$0.301mGy. $\cdot$ The percentage of the absorber finding from phantom was more the 85% with a range of 100$\sim$136 kvp. $\cdot$ The mean dose from phantom was 0.990mGy with 124 kvp and 16 mAs. $\cdot$ There is no difference in the amount of information due to the change of kvp and mAs, but some coarseness of images can be seen in the magnified images. $\cdot$ The resolving power was about 1.5/mm. $\cdot$ The increase and decrease of 10$\sim$30 kvp from common kvp has not influenced on the amount of information. $\cdot$ The exposure could be reduced by 0.170$\sim$0.570mGy.