The Correction of Fluid Temperature for Hot-wire Anemometer

열선 유속계에 대한 유체 온도의 보정

  • 심상학 (고창기능대학 메카트로닉스과)
  • Published : 1999.12.01

Abstract

This paper reports by simple method that is quickly corrected the effects of fluid temperature for the hot wire anemometer. We are concerned with a variable output of hot wire anemometer on arbitrary fluid temperature. Hot wire by measuring boundary layer of turbulent flow has been calibrated by arbitrary temperature lower than 10$0^{\circ}C$, and velocity lower than 20m/s. As a result, we could pick up the temperature factor affected by output of hot wire anemometer from related in output of arbitrary temperature to output of room temperature. By using temperature factor on the output of hot wire anemometer, we also obtained that the relationship of velocity was of no effect by temperature of fluids. About results of calibrated hot wire, uncertainly of velocity is 2.15% at room temperature and 3.1% at arbitrary temperature.

Keywords

References

  1. 熱線流速計 v.1;2;3;4;5;6 流れの計測(連載) 西岡通南
  2. Turbulence(2nd ed.) J. O.Hinze
  3. J. Fluid Mech. v.76 A note on static and dynamic calibration of constant-temperature hot-wire probes H. H. Bruun
  4. J. Fluid Mech. v.6 Two-dimensional convection from heated wires at the low Reynolds numbers D. C. Collis;M. J. Williams
  5. Experiments in Fluids v.13 On the effects of fluid temperature on hot wire characteristic(Part 1: Results of experiments) M. Kostka;R. V. Vasanta
  6. Experiments in Fluids v.13 On the effects of fluid temperature on hot wire characteristic(Part 2: Foundation of a rational theory) R. V. Vasanta
  7. 日本機械學會論文集(B編) v.64 no.620 發達チャンネル亂流の亂れ特性に及ばすレイノルズ數の效果 金宰基;吉識晴夫
  8. 日本機械學會論文集(B編) v.64 no.627 發達チャンネルにおける壁面加熱の影響 金宰基;吉識晴夫
  9. ASME性能試驗規約「計測の不確かさ」 日本機械學會(譯)