Comparison of Precision for Available Phosphate Measurement between Lancaster and Soiltek KA-1 Spectrophotometric Method

Lancaster 법과 Soiltek KA-1 분광광도계를 이용한 유효인산 측정법의 정밀도 비교

  • Published : 1999.09.30

Abstract

To measure the precision of Soiltek$^{(R)}$ KA-1 spectrophotometric measurement of soil available phosphate, standard deviation, C;V, and relative sensitivity were investigated, comparing to Lancaster method. Available phosphate concentration measured by Soiltek$^{(R)}$ KA-1 spectrophotometric method was highly correlated with Lancaster method($r=0.998^{***}$) and the slop of regression equation was close to 1.0. There were no differences in standard deviation. CV, and relative sensitivity for soil available phosphate concentrations measured by two methods. Regarding to convenience of experimental procedure and instrumental operation, and time saving analysis, Soiltek$^{(R)}$ KA-1 spectrophotometric measurement for soil available phosphate is highly recommended.

토지이용 유형과 유효인산 함량이 다양한 56개 토양을 대상으로 범용 분광광도계를 사용한 Lancaster법과 분석과정 및 측정과정을 간편화시킨 Soiltek$^{(R)}$ KA-1에서 제시하는 방법으로 측정한 유효인산함량에 대하여 정확도 및 정밀도를 비교 분석하였다. 두 방법으로 분석된 인산함량은 고도의 유의성 있는 정의상관 ($r=0.998^{***}$)을 보였으며, 기울기는 0.97로 거의 일치하여 두 방법간의 정확성은 차이가 없었다. 또한, 두 방법으로 측정된 인산함량 분석결과의 표준편차, CV, 상대정밀도(RS)는 거의 차이가 없으므로 정밀도에서도 차이가 없었다. 따라서, 분석과정 및 기기조작의 편리성 및 신속성을 고려한다면 Soiltek$^{(R)}$ KA-1에서 제시하는 방법으로 유효인산 함량을 측정하는 것도 유용한 방법이라고 생각된다.