MEASUREMENT OF PULPAL BLOOD FLOW USING A LASER DOPPLER FLOWMETER

Laser Doppler flowmeter를 이용한 치수혈류 측정

  • Ban, Tae-Whan (Department of Conservative Dentistry, College of Dentistry, Kyungpook National University) ;
  • Lee, Jae-Sang (Department of Conservative Dentistry, College of Dentistry, Kyungpook National University) ;
  • Kim, Sung-Kyo (Department of Conservative Dentistry, College of Dentistry, Kyungpook National University)
  • 반태환 (경북대학교 치과대학 치과보존학교실) ;
  • 이재상 (경북대학교 치과대학 치과보존학교실) ;
  • 김성교 (경북대학교 치과대학 치과보존학교실)
  • Published : 1999.12.02

Abstract

Blood supply rather than nerve supply implies pulp vitality. To evaluate pulp vitality clinically, electric pulp test and thermal test which are based on sensory nerve response have been used in addition to many auxiliary data such as past dental history, visual inspection, radiographic examination, percussion, palpation and transillumination test. However, reactivity of the nerves to the stimulation is not synonymous with normalcy. Therefore measurement of pulpal blood flow using a laser Doppler flowmeter became a new trial to test the pulp vitality. The purpose of the present study was to evaluate normal pulpal blood flow level of maxillary teeth in adult to provide a guideline in determining the vitality of dental pulp. Pulpal blood flow was measured in maxillary central and lateral incisors, canines, first and second premolars and first molars of seventy nine adults of 22 - 30 years old using a laser Doppler flowmeter (PeriFlux 4001, Perimed Co., Stockholm, Sweden, 780 nm infrared laser, 1mW). For directly-made splints, silicone rubber impressions were taken directly from the mouth. For indirectly-made splints, alginate impressions were taken from the mouth and stone cast were made. After making depressions on the buccal surfaces of the cast teeth to indicate the hole positions, second impressions with vinyl polysyloxane putty were taken from the cast. Holes for the laser probes were made at the putty impressions 4mm above the gingival level. Laser probe (PF416 dental probe, 1.5mm) was inserted in the prepared hole and the splint was set in the mouth. After 10 minutes of patient relaxing, pulpal blood flow was recorded for 5 minutes on each tooth. The recorded flow was saved in the computer and calculated with a software 'Perisoft' version 5.1. Pulpal blood flow was also recorded in six teeth of five individuals with no response to electric pulp test and cold test, with periapical radiolucency, or with history of root canal treatment to compare with nonvital teeth. The difference between the mean flow values of each group of teeth were analyzed using one-way ANOVA and Duncan's Multiple Range test. The results were as follows: 1. The average pulpal blood flow values of all the tested teeth of each location were between 9 - 16 Perfusion Unit. Pulpal blood flow value was highest in maxillary lateral incisors, followed by first premolars, second premolars, canines, central incisors, and then first molars (p<0.01). 2. In six anterior teeth, indirectly-made splint group showed higher pulpal blood flow values than directly-made splint group (p<0.01). In posterior teeth, however, there was no significant flow value difference between directly-made splint group and indirectly-made splint one (p>0.05). 3. Teeth with vital pulps showed higher signal values than teeth with nonvital pulps (p<0.01), and the flow photographs showed heartbeat-synchronous fluctuations and vasomotions, while those were absent in non vital tooth.

치수생활력이란 치수내의 혈관의 상태를 가리킨다. 치수생활력 측정을 위해서 사용되어져 온 온도검사나 전기치수 검사는 치수신경의 반응에 기초하여 간접적으로 치수의 생활력을 판단하는 방법이다. 따라서 치수생활력의 판단을 위해서 치수생활력의 기본인 치수혈류를 파악할 필요가 있어 laser Doppler flowmeter를 이용하여 치수 혈류를 측정하는 방법이 연구되었다. 본 연구에서는 성인치아에서 치수생활력 검사를 위해 laser Doppler flowmeter를 이용하여 치수혈류를 측정 할 경우 생활력의 판단 근거로 참고할 수 있는 정상혈류 측정치의 기초자료를 마련할 목적으로, laser Doppler flowmeter를 이용하여 성인의 상악 중절치에서부터 제1대구치에 이르기까지 각 정상 치아의 치수혈류를 측정하고, probe 고정용 splint 제작 방법 및 생활치수와 실활치수에 따른 치수혈류 측정치의 차이를 비교, 연구하였다. 22세에서 30세까지 성인 남녀 79명을 대상으로 상악 중 절치에서부터 제1대구치에 이르기까지 laser Doppler flowmeter를 사용하여 치수혈류를 측정하였고, 생활치아에 대한 비교군으로는 실활치아라고 판단되는 치아를 가진 성인 남녀 자원자 5명의 치아 13개도 함께 검사하였다. 혈류 측정 probe고정용 splint제작을 위한 고무인상시 직접법에서는 환자 구강내에서 직접 고무인상을 채득하였고, 간접법에서는 구강내에서 일차 인상을 채득하여 만든 석고모형상에서 probe 위치를 표시한 다음 석고모형상에서 이차적으로 고무인상을 채득하였다. 고무인상에서 치관부에 probe 고정을 위한 hole을 형성하고 probe를 삽입, 고정한 다음 probe를 치아 협면에 적용하였다. Laser Doppler flowmeter에서 안정된 신호가 나타날 때까지 피검자를 10분 이상 동안 안정시켰으며 안정상태 도달 후 각 치아마다 5 분간 지수혈류를 기록하였다. 측정된 치수혈류는 'Perisoft'(Perimed Co., Sweden) 프로그램을 이용하여 컴퓨터에 저장한 후 측정치를 계측하였다. 각 치아간의 실험치 평균의 차이는 일원 변량 분석법(one-way ANOVA) 및 Duncan's Multiple Range test를 이용하여 분석하였으며, 다음과 같은 결과를 얻었다. 1. 상악 중절치에서부터 제1대구치에 이르기까지 시험대상 전 치아에서의 치수혈류 측정치 평균은 9 - 16 Perfusion Unit 사이로 나타났는데, 상악 측절치에서 가장 높게 나타났고, 제1소구치, 제2소구치, 견치, 중절치, 제1대구치의 순으로 높게 나타났다(p<0.01). 2. 상악 6전치에서는 간접인상법으로 제작된 splint 적용군이 직접인상법으로 제작된 splint 적용군에 비해 높은 측정치를 나타내었으나(p<0.01). 소구치 및 대구치에서는 측정치에 있어서 차이를 나타내지 않았다(p>0.05) 3. 생활치아에서는 실활치아에 비해 현저히 높은 측정치를 나타내었고(p<0.01). 측정선의 양상은 심장 박동과 일치하는 파동성 및 율동성을 나타내었다.

Keywords