Measurements of Setup Error and Physiological Movement of Liver by Using Electronic Portal Imaging Device in Patients with Hepatocellular Carcinoma

간암환자에서 Electronic Portal Imaging Device(EPID)를 이용한 자세 오차 및 종양 이동 거리의 객관적 측정

  • Keum Ki Chang (Department of Radiation Oncology, Yonsei University, College of Medicine, Yonsei Cancer Center) ;
  • Lee Sang-wook (Department of Radiation Oncology, Yonsei University, College of Medicine, Yonsei Cancer Center) ;
  • Shin Hyun Soo (Department of Radiation Oncology, Yonsei University, College of Medicine, Yonsei Cancer Center) ;
  • Kim Gwi Eon (Department of Radiation Oncology, Yonsei University, College of Medicine, Yonsei Cancer Center) ;
  • Sung Jinsil Seong (Department of Radiation Oncology, Yonsei University, College of Medicine, Yonsei Cancer Center) ;
  • Lee Chang Geol (Department of Radiation Oncology, Yonsei University, College of Medicine, Yonsei Cancer Center) ;
  • Chu Sung Sil (Department of Radiation Oncology, Yonsei University, College of Medicine, Yonsei Cancer Center) ;
  • Chang Sei-Kyung (Department of Radiation Oncology, Yonsei University, College of Medicine, Yonsei Cancer Center) ;
  • Suh Chang Ok (Department of Radiation Oncology, Yonsei University, College of Medicine, Yonsei Cancer Center)
  • 금기창 (연세대학교 의과대학 치료방사선과학교실, 연세암센터) ;
  • 이상욱 (연세대학교 의과대학 치료방사선과학교실, 연세암센터) ;
  • 신현수 (연세대학교 의과대학 치료방사선과학교실, 연세암센터) ;
  • 김귀언 (연세대학교 의과대학 치료방사선과학교실, 연세암센터) ;
  • 성진실 (연세대학교 의과대학 치료방사선과학교실, 연세암센터) ;
  • 이창걸 (연세대학교 의과대학 치료방사선과학교실, 연세암센터) ;
  • 추성실 (연세대학교 의과대학 치료방사선과학교실, 연세암센터) ;
  • 장세경 (연세대학교 의과대학 치료방사선과학교실, 연세암센터) ;
  • 서창옥 (연세대학교 의과대학 치료방사선과학교실, 연세암센터)
  • Published : 2000.06.01

Abstract

Purpose : The goal of this study 닌as to improve the accuracy of three-dimensional conformal radiotherapy (3-D CRT) by measuring the treatment setup error and physiological movement of liver based on the analysis of images which were obtained by electronic portal imaging device (EPID). Materials and Methods : For 10 patients with hepatocellular carcinoma, 4-7 portal images were obtained by using EPID during the radiotherapy from each patient daiiy. We analyzed the setup error and physiological movement of liver based on the verification data. We also determined the safety margin of the tumor in 3-D CRT through the analysis of physiological movement. Results : The setup errors were measured as 3mm with standard deviation 1.70 mm in x direction and 3.7 mm with standard deviation 1.88 mm in y direction respectively. Hence, deviation were smaller than 5mm from the center of each axis. The measured range of liver movement due to the physiological motion was 8.63 mm on the average. Considering the motion of liver and setup error, the safety margin of tumor was at least 15 mm. Conclusion : EPID is a very useful device for the determination of the optimal margin of the tumor, and thus enhance the accuracy and stability of the 3-D CRT in patients with hepatocellular carcinoma.

목적 : 간암의 3차원 입체조형치료계획시 electronic portal imaging device (EPID)를 이용한 검증영상으로 경계여유 결정에 필요한 객관적 기초자료를 얻으므로써 방사선치료의 정확성을 높이고 본 기기를 이용하여 방사선치료의 정도관리 측면에서의 유용성을 알아보고자 하였다. 대상 및 방법: 10명의 간암 환자에서 EPID를 이용하여 1회 방사선 치료 중 약 4극회의 검증영상을 획득하였고 총 10회 분할 치료기간 동안 반복하여 시행하였다. 방사선치료 자세 고정시 발생할 수 있는 각 개인의 자세 오차를 구하여 치료간 발생하는 환자의 위치 이동을 분석하였고, 호흡으로 인한 간의 상하 움직임의 정도를 측정하여 계획용표적체적(Planning Target Volume, PTV)의 결정시 필요한 적정 경계 여유를 구하였다. 결과 : 검증영상 분석 결과 간암의 치료시 자세이동에 의한 자세 오차는 x축으로 3.0 mm, 표준편차는 1.70 mm 였고 y축으로 3.7 mm, 표준편차 1.88 mm 였다. 따라서 각 축을 중심으로 5 mm 이내였다. 또한 호흡에 의한 간의 상하 운동 범위는 평균 8.63 mm이었고 표준편차는 1.48 mm 였다. 따라서 계획용표적체적 결정시 종양에 적어도 15 mm의 경계여유가 더 필요하다 결론 : EPID는 3차원 입체조형치료계획시 종양의 경계여유 결정에 매우 유용하며 나아가서 방사선치료의 정도관리에도 크게 기여할 것이다