Analysis of the Inter- and Intra-treatment Isocenter Deviations in Pelvic Radiotherapy With Small Bowel Displacement System
Small Bowel Displacement System을 이용한 골반부 방사선조사에서 치료간 및 치료중 중심점 위치변동에 관한 분석
- Kim Moon Kyung (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
- Kim Dae Yong (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
- Ahn Yong Chan (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
- Huh Seung Jae (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
- Lim Do Hun (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
- Shin Kyung Hwan (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
- Lee Kyu Chan (Department of Radiation Oncology, Gil Medical Center, Gachon Medical School)
- 김문경 (성균관대학교 의과대학 삼성서울병원 치료방사선과) ;
- 김대용 (성균관대학교 의과대학 삼성서울병원 치료방사선과) ;
- 안용찬 (성균관대학교 의과대학 삼성서울병원 치료방사선과) ;
- 허승재 (성균관대학교 의과대학 삼성서울병원 치료방사선과) ;
- 임도훈 (성균관대학교 의과대학 삼성서울병원 치료방사선과) ;
- 신경환 (성균관대학교 의과대학 삼성서울병원 치료방사선과) ;
- 이규찬 (가천 의과대학 부속 길병원 방사선종양학과)
- Published : 2000.06.01
Abstract
Purpose : To evaluate the e지ent and frequency of the inter- and intra-treatment isocenter deviations of the whole pelvis radiation field in using small bowel displacement system (SBDS). Methods and Materials : Using electronic portal imaging device (EPID), 302 postero-anterior 232 lateral portal images were prospectively collected from 11 patients who received pelvic radiation therapy (7 with cervix cancer and 4 with rectal cancer). All patients were treated in prone position with SBDS under the lower abdomen. Five metallic fiducial markers were placed on the image detection unit for the recognition of the isocenter and magnification. After aligning the bony landmarks of the EPID images on those of the reference image, the deviations of the isocenter were measured in right-left (RL), cranio-caudal (CC), and PA directions. Results : The mean inter-treatment deviation of the isocenter in each RL, CC, and PA direction was 1.2 mm (
목적 : small bowel displacement system (SBDS)을 이용한 골반부 방사선 치료시 치료간(intertreatment) 및 치료중(intratreatment) 중심점 위치변동의 편차를 분석하고, 그 결과를 임상적으로 이용하고자 하였다. 대상 및 방법 :자궁경부암 7명, 직장암 4명의 총 11명의 환자를 대상으로 electronic portal imaging device (EPID)를 이용하여 302건의 후-전문(postero-anterior port) 영상과 232건의 측문(lateral port) 영상을 얻었다. 모든 환자는 복와위 자세로 하복부에 SBDS를 사용하였다 방사선 조사야의 중심점과 배율을 맞추기 위하여 5개의 금속 기준점을 영상탐지판(image detection unit)에 부착시켰으며, EPID 영상의 골 기준점(bony landmark)을 정하여 모의치료시 촬영한 영상과 비교하여 정렬시킨 후 우-좌문, 두-미문, 그리고 후-전문 방향으로 중심점의 이동방향 및 거리를 분석하였다. 결과 : 우-좌문, 두-미문, 그리고 후-전문 방향으로의 치료간 중심점 이동의 평균값은 각각 1.2 mm (