Radiation Oncology Journal
- Volume 19 Issue 1
- /
- Pages.66-73
- /
- 2001
- /
- 2234-1900(pISSN)
- /
- 2234-3156(eISSN)
Dosimetry of the Low Fluence Fast Neutron Beams for Boron Neutron Capture Therapy
붕소-중성자 포획치료를 위한 미세 속중성자 선량 특성 연구
- Lee, Dong-Han (Laboratory. of Radiation Effect, Korea Cancer Center Hospital) ;
- Ji, Young-Hoon (Laboratory. of Radiation Effect, Korea Cancer Center Hospital) ;
- Lee, Dong-Hoon (Laboratory. of Radiation Effect, Korea Cancer Center Hospital) ;
- Park, Hyun-Joo (Laboratory. of Radiation Effect, Korea Cancer Center Hospital) ;
- Lee, Suk (Laboratory. of Radiation Effect, Korea Cancer Center Hospital) ;
- Lee, Kyung-Hoo (Laboratory. of Radiation Effect, Korea Cancer Center Hospital) ;
- Suh, So-Heigh (Laboratory. of Radiation Effect, Korea Cancer Center Hospital) ;
- Kim, Mi-Sook (Department of Radiation Oncology, Korea Cancer Center Hospital) ;
- Cho, Chul-Koo (Department of Radiation Oncology, Korea Cancer Center Hospital) ;
- Yoo, Seong-Yul (Department of Radiation Oncology, Korea Cancer Center Hospital) ;
- Yu, Hyung-Jun (Department of Radiation Oncology, Korea Cancer Center Hospital) ;
- Gwak, Ho-Shin (Department of Neurosurgery, Korea Cancer Center Hospital) ;
- Rhee, Chang-Hun (Department of Neurosurgery, Korea Cancer Center Hospital)
- 이동한 (원자력병원 방사선영향연구실) ;
- 지영훈 (원자력병원 방사선영향연구실) ;
- 이동훈 (원자력병원 방사선영향연구실) ;
- 박현주 (원자력병원 방사선영향연구실) ;
- 이석 (원자력병원 방사선영향연구실) ;
- 이경후 (원자력병원 방사선영향연구실) ;
- 서소희 (원자력병원 방사선영향연구실) ;
- 김미숙 (원자력병원 방사선종양학과) ;
- 조철구 (원자력병원 방사선종양학과) ;
- 류성렬 (원자력병원 방사선종양학과) ;
- 유형준 (원자력병원 방사선종양학과) ;
- 곽호신 (원자력병원 신경외과) ;
- 이창훈 (원자력병원 신경외과)
- Published : 2001.03.01
Abstract
Purpose : For the research of Boron Neutron Capture Therapy (BNCT), fast neutrons generated from the MC-50 cyclotron with maximum energy of 34.4 MeV in Korea Cancer Center Hospital were moderated by 70 cm paraffin and then the dose characteristics were investigated. Using these results, we hope to establish the protocol about dose measurement of epi-thermal neutron, to make a basis of dose characteristic of epi-thermal neutron emitted from nuclear reactor, and to find feasibility about accelerator-based BNCT. Method and Materials : For measuring the absorbed dose and dose distribution of fast neutron beams, we used Unidos 10005 (PTW, Germany) electrometer and IC-17 (Far West, USA), IC-18, ElC-1 ion chambers manufactured by A-150 plastic and used IC-l7M ion chamber manufactured by magnesium for gamma dose. There chambers were flushed with tissue equivalent gas and argon gas and then the flow rate was S co per minute. Using Monte Carlo N-Particle (MCNP) code, transport program in mixed field with neutron, photon, electron, two dimensional dose and energy fluence distribution was calculated and there results were compared with measured results. Results : The absorbed dose of fast neutron beams was
목적 : 붕소-중성자 포획치료법(Boron Neutron Capture Therapy, BNCT)을 위해 원자력병원 싸이클로트론에서 발생되는 최대에너지 34.4 MeV의 속중성자(Fast neutron)를 70 cm 파라핀으로 감속시킨 후 선량 특성을 조사하였다. 그 결과를 토대로 열외중성자(Epithermal neutron) 선량 측정법에 대한 프로토콜을 확립하여 원자로에서 방출되는 열외 중성자 선량 특성 평가의 기초를 삼고, 가속기를 이용한 BNCT 연구에 대한 타당성 여부를 조사하고자 한다. 대상 및 방법 : 공기 중 선량 및 물질 내 선량 분포 측정을 위해 Unidos 10005 (PTW, Germany) 전기계와 조직 등가 물질인 A-150 플라스틱으로 제작된 IC-17 (Far West, USA) 및 IC-18, ElC-1 이온함을 사용하였고, 감마선의 측정을 위해서는 마그네슘으로 제작된 IC-l7M 이온함을 이용하였으며 조직등가 기체와 아르곤 기체를 분당 5cc 씩 주입하며 측정하였다. 중성자, 광자, 전자가 혼합된 장의 모의 수송 해석을 위해 이용되는 Monte Carlo N-Particle (MCNP) transport code를 사용하여 2차원적 선량 분포 및 에너지 분포를 계산하였으며 이 결과를 측정값과 비교하였다. 결과 : BNCT에서의 유효 치료 깊이인 물 팬텀 4 cm에서의 선량은 치료기 1 MU 당
Keywords