The Role of Air-Vacuum Cushion Device in Patients with Rectal Cancer in Radiation Therapy

직장암 환자에서 방사선치료시 Air-vacuum Cushion의 유용성

  • Kim Ki-Hwan (Department of Therapeutic Radiology, Chungnam National University Hospital) ;
  • Cho Moon-June (Department of Therapeutic Radiology, Chungnam National University Hospital, Cancer Research Institute, Chungnam National University) ;
  • Kang No-Hyun (Department of Therapeutic Radiology, Chungnam National University Hospital) ;
  • Kim Dong-Wuk (Department of Therapeutic Radiology, Chungnam National University Hospital) ;
  • Kim Jun-Sang (Department of Therapeutic Radiology, Chungnam National University Hospital, Cancer Research Institute, Chungnam National University) ;
  • Jang Ji-Young (Department of Therapeutic Radiology, Chungnam National University Hospital) ;
  • Kim Jae-Sung (Department of Therapeutic Radiology, Chungnam National University Hospital, Cancer Research Institute, Chungnam National University)
  • 김기환 (충남대학교 의과대학 치료방사선과학교실) ;
  • 조문준 (충남대학교 의과대학 치료방사선과학교실, 충남대학교 암공동연구소) ;
  • 강노현 (충남대학교 의과대학 치료방사선과학교실) ;
  • 김동욱 (충남대학교 의과대학 치료방사선과학교실) ;
  • 김준상 (충남대학교 의과대학 치료방사선과학교실, 충남대학교 암공동연구소) ;
  • 장지영 (충남대학교 의과대학 치료방사선과학교실) ;
  • 김재성 (충남대학교 의과대학 치료방사선과학교실, 충남대학교 암공동연구소)
  • Published : 2001.09.01

Abstract

Prupose : We analyzed setup errors induced by using air-vacuum cushion as immobilization device in patients with rectal cancer. Materials and methods : We had treated the twenty patients with rectal cancer by 6 MV, 10 MV X-ray from Aug. 1998 to Aug. 1999 at Chungnam National University Hospital. All patients were treated at prone position. They were separated to two groups, control group, 10 patients using styrofoam, and test group, 10 patients using styrofoam and air-vacuum cushion. We measured errors of posterior field for x, y axis and lateral field for z, y axis with simulation film and EPID image using a matching technique. Results : In control group, the mean displacement values of pelvic bone landmark for x axis and y axis were 0.02 mm. 0.78 mm, respectively and the standard deviations of systematic error were 2.13 mm, 2.40 mm, respectively and the standard deviation of random error were 1.46 mm. 1.51 mm, respectively. In test group, the mean displacement values of x axis and y axis were -0.33 mm. 0.81 mm, respectively and the standard deviations of systematic error were 1.71 mm, 3.08 mm, respectively and the standard deviations of random errors were 1.40 mm. 1.88 mm, respectively. The mean displacement values of z axis and y axis were 2.98 mm. 0.74 mm, respectively and the standard deviations of systematic error were 4.75 mm, 2.65 mm, respectively and standard deviations of random error were 2.69 mm. 1.86 mm, respectively. The statistical difference of field size by using air vacuum cushion between two groups in posterior direction and lateral direction was not shown. Conclusion : We think that use of air-vacuum cushion may not be an advantage for improving setup accuracy in rectal cancer patients.

목적 : 본 연구에서는 골반부위 방사선치료 환자중 직장암 환자에서 고정기구로 사용한 air-vacuum cushion의 조사면의 변화에 따른 고정기구의 유용성을 후향적으로 분석하였다. 대상 및 방법 : 충남대학교병원 치료방사선과에서 1998년 8월부터 1999년 8월까지 직장암으로 방사선치료를 받은 20명의 환자를 대상으로 하였다. 환자들은 모두 복와위 자세로 치료를 받았고 대조군 10명과 실험군 10명으로 각각 나누어서 연구를 실행하였다. 전체 치료과정중 EPID를 이용하여 후면조사면은 x축(right-left), y축(cranio-caudal)에 대한 영상을 얻었고, 측면조사면은 z축(anterior-posterior), y축(cranio-caudal)에 대한 영상을 얻었다. 전체 치료과정 중 얻은 영상들을 모의치료시 촬영한 기준필름과 matching 기법을 이용하여 오차를 측정하였다. 통계적 유의성 확인은 t-검정법을 사용하였다. 결과 : 대조군에서 10명 환자의 후면 조사영역의 경우 x축, y축에 대한 평균이동값은 각각 0.02 mm, 0.78 mm이었고 계통오차과 우연오차의 표준편차값은 각각 2.13 mm, 2.40 mm와 1.46 mm, 1.51 mm이었다. 측면조사면에서 z축, y축에 대한 평균이동값은 각각 2.96 mm, 0.47 mm이었고 계통오차와 우연오차의 표준편차는 각각 3.69 mm, 1.96 mm와 2.79 mm, 1.48 mm이었다. 실험군 10명 환자의 후면 조사면에서 x축, y축에 대한 평균이동값은 각각 -0.33 mm, 0.81 mm이었고 계통오차과 우연오차에 대한 표준편차는 각각 1.71 mm, 3.08 mm와 1.40 mm, 1.88 mm이었다. 측면조사면에서 z축, y축에 대한 평균이동값은 각각 2.98 mm, 0.74 mm이었고 계통오차과 우연오차에 대한 표준편차는 각각 4.75 mm, 2.65 mm와 2.69 mm, 1.86 mm이었다. 본 연구에서는 대조군과 실험군간의 조사면에 대한 평균 이동값과 우연오차는 거의 일치하였으나 계통오차는 실험군에서 다소 큰 경향을 보였다. 후면조사면과 측면조사면에서 air-vacuum cushion에 의한 조사면 변화의 통계적 유의성이 없었다. 결론 : 직장암 환자의 방사선 치료시 고정기구로 사용한 air-vacuum cushion의 사용이 방사선치료의 정확도에 이익을 주지 않는 것으로 생각된다.

Keywords