Development of pub Level Dissolved Oxygen Measuring Technology using Clark Cell

Clark Cell을 이용한 ppb단위 용존산소 측정기술개발

  • 정경열 (한국기계연구원 환경설비 연구부) ;
  • 류길수 (한국해양대학교 자동화 정보공학부)
  • Published : 2002.05.01

Abstract

Measuring of the dissolved oxygen is widely used for the environmental control of natural waters, sewage waste treatment, medical and biochemical studies, soil husbandry, food and drug process control, and prevention of corrosion in boilers. Especially, a power plants need special management for preventing accidents from corrosion, therefore, it is essential to measure the concentration of dissolved oxygen in real-time. In this paper we present a method of measuring dissolved oxygen very accurately up to pub units. This method, called polarograph method, is based on the measures of the electric current generated by the oxidation process in cathode and de-oxidation process in anode, assuming that the amount of the current is proportional to the density of dissolved oxygen.

Keywords

References

  1. 한국표준연구원 제출보고서 DO Meter 개발에 관한 연구 이광우 (외3인)
  2. Measurement of Dissoloved Oxygen M.L. Hitchman
  3. Design of Membrane-Covered Polarographic Gas Detectors v.41 D.P. Lucero
  4. Sensors and Actuators v.B21 Disposable Clark oxygen electrode using recycled materials and its application H. Suzuki
  5. OP앰프의 실무회로 이병식
  6. 입문에서 응용까지 OP 앰프 실전기술 谷本 ?
  7. C로 쓰는 8051 박귀태;이상락
  8. 8051 마이크로프로세서-Embedded Microcontroller 8051 한홍석(외 4인)
  9. http://www. semiconductors.philips.com/mcu/app-lications/
  10. 3600 Analysis for Oxygen, Orbisphere 3600 model Manual
  11. Rosemount Analytical model 7001D Oxysgen monitors, Rosemount 7001D model Manual