A Study on Quality Assurance for Gamma Knife

감마나이프 정도관리에 관한 연구

  • 서원섭 (경희대학교 의과대학 신경외과학교실) ;
  • 신동오 (경희대학교 의과대학 방사선종양학교실) ;
  • 지영훈 (원자력의학원 방사선종양학과) ;
  • 임영진 (경희대학교 의과대학 신경외과학교실)
  • Published : 2003.09.01

Abstract

It is strongly recommended that periodic quality assurance should be carried out in the Gamma Knife that is used in radiosurgery since high radiation is delivered in one session. Since the protocols for Gamma Knife recommended by associations or agencies on quality assurance are absent in Korea, hospitals possessing the Gamma Knife have developed their own protocols. In order to develop a quality assurance protocol suitable for Korea, we reviewed the protocols of the Gamma Knife manufacturer, USA and Japan. we categorized the periodic items into three parts,: radiation dose, mechanical and safety part. The USA recommended and regulated more strict than the manufacturer. Japan recommended the items and frequency based on the USA. In conclusion, we tried to suggest a basic Gamma Knife quality assurance protocol suitable for Korea.

감마나이프는 일회에 대용량의 방사선량을 조사하여 뇌정위 방사선 수술을 하기 때문에, 주기적인 정도 관리가 매우 중요하다. 그러나 국내의 경우 학회 및 규제기관 차원에서 권고되고 있는 표준화된 정도관리 절차서가 없이, 각 기관별로 독자적인 정도관리 절차서를 개발하여 시행되고 있어, 기관별로 상호 비교하기가 어려운 실정이다. 따라서 본 연구에서는 국내 실정에 적합한 감마나이프 정도관리 프로그램을 개발하기 위해, 제조사, 미국 및 일본에서 권고하고 있는 정도관리 절차서를 수집하여 비교 분석하였다 또한 감마나이프에 대한 주기적인 정도관리 검사항목을 선량관련 검사항목, 기계적인 검사항목 및 안전 관리적인 측면으로 분류하였다. 미국의 경우 제조사 보다 엄격하게 권고 및 규제하고 있으며, 일본의 경우 미국의 절차서를 토대로 검사항목 및 주기를 권고하고 있다. 이를 토대로 국내 실정에 적합한 감마나이프 정도관리 절차를 제시하고자 한다.