Uncertainties in Pressure Calibration of Laboratory Standard Microphones by Reciprocity Technique

가역방법에 의한 표준 마이크로폰 음압교정의 불확도

  • 서상준 (한국표준과학연구원 물리표준부 음향진동그룹) ;
  • 권휴상 (한국표준과학연구원 물리표준부 음향진동그룹) ;
  • 이용봉 (한국표준과학연구원 물리표준부 음향진동그룹) ;
  • 서재갑 (한국표준과학연구원 물리표준부 음향진동그룹)
  • Published : 2004.02.01

Abstract

According to the Mutual Recognition Arrangement (MRA), the calibration and/or test laboratories should satisfy the management and technical requirements ISO 17025 or equivalent. Chapter 5, Section 5.10.4 of the technical requirement of ISO 17025 suggests the required informations for calibration certificates, one of them is to state the uncertainty of measurement. The uncertainties of measurement in reciprocity calibration of standard laboratory microphone were calculated. The expanded uncertainties for 1 and 1/2 inch microphones were 0.03 dB in the middle frequency range and they increased up to 0.10 dB and 0.11 dB at 20 Hz, 0.07 dB and 0.08 dB at high frequency, respectively.

교정 및/또는 시험기관은 국제상호인증협정 (Global Mutual Recognition Arrangement)에 따라 ISO 17025[1]의 경영요건 및 기술요건의 제반 조건 또는 이와 동등한 조건을 충족시켜야 한다. ISO 17025의 기술요건을 기술한 제 5장, 제 5.10.4 절은 교정성적서 (Calibration Certificate)에 대한 요구사항을 제시하고 있는데, 이들 중 하나가 측정 불확도 (Uncertainty of Measurement)를 교정성적서에 명시하여야 하는 것이다. 본 논문에서는 표준 마이크로폰의 가역교정에 따른 제반 불확도 요인을 측정 또는 계산하여 표준 마이크로폰의 교정에 대한 측정 불확도를 계산하였다. 중간 주파수 영역에서 1인치 및 1/2 인치 마이크로폰의 경우 확장불확도는 0.03 dB 였으며 20 Hz에서는 각각 0.10 dB과 0.11 dB, 고주파에서는 각각 0.07 dB와 0.08 dB까지 증가한다.

Keywords

References

  1. ISO 17025 General Requirements for the Competance of Testing and Calibration Laboratories
  2. Guide to the Expression of Uncertainty in Measurement ISO
  3. IEC 61094-1 Measurement Microphones - Part 1, Specifications for Laboratory Standard Microphones
  4. IEC 61094-2 Measurement Microphones - Part 2, Primary Method for Pressure Calibration of Laboratory Standard Microphones by the Reciprocity Technique
  5. B&K Technical Documents of Reciprocity Calibration System Type 9699
  6. Acta Acustika v.1 Radial wave-motion in cylindrical plane-wave couplers Rasmussen,K.
  7. 측정불확도 표현 지침, KRISS-99-070-SP 한국표준과학연구원
  8. Instruction Manual of B&K 4160 and 4180
  9. Metrologia v.36 The static pressure and temperature coefficients of laboratory standard microphones Rasmussen,K.
  10. B&K Technical Review no.1 The influence of environmental conditions on the pressure sensitivity of measurement microphones Rasmussen,K.
  11. Metrologia v.36 Controlled environment for reciprocity calibration of laboratory standard microphones and measurement of sensitivity pressure correction Wong,G.S.K.;Wu,L.