Abstract

This paper presents the vibration analysis and the performance evaluation of cantilevers with probing tips for non-contact scanning probe microscopy. One of the current issues of the scanning probe microscopy technology is to increase the measurement speed, which is closely tied with the dynamic characteristics of cantilevers. The primary concern in this research is to investigate the relation between the maximum possible speed of non-contact scanning probe microscopy and the dynamic characteristics of cantilevers. First, the finite element analysis is made for the vibration characteristics of various cantilevers in use. The computed natural frequencies of the cantilevers are in good agreement with measured ones. Then, each cantilever is tested with topographic measurement for a standard sample with the scanning speed changed. The performances of cantilevers are analyzed along with the natural frequencies of cantilevers. Experiments are also performed to test the effects of how to attach cantilevers in the piezo-electric actuator. Finally, measurement sensitivity has been analyzed to enhance the performance of scanning probe microscopy.

Keywords

References

  1. Sarid, D., 1991, Scanning Force Microscopy: with applicarions to electric, Magnetic, and Atomic forces, New York Oxford University Press.
  2. 2000, User's Guide to Autoprobe M5, TM Microscopes
  3. Park, S. K., Kim, Y. H. and Ha, S. K, 2002, "Analysis of Dynamic Behavior of Piezoelectric Atomic Force Microscope Cantilever," Journal of the Korean Society for Noise and Vibration Engineering, Vol. 12, No. 3, pp. 187-194. https://doi.org/10.5050/KSNVN.2002.12.3.187
  4. Chang, W. S., Shin, B. S. and Whang, K. H., 2003, "Nanoprobe Application Technologies," Journal of the Korean Society of Precision Engineering, Vol. 20, No. 3, pp. 5-14.
  5. Park, S., Hong, I. J. and Nho, Y., 2003, "Scanning probe microscopy," http://www.psia.co.kr.
  6. Binning, G. and Quate, C. F., 1986, "Atomic Force Microscope," Phys. Rev. Lett., Vol. 56(9), pp. 930-933. https://doi.org/10.1103/PhysRevLett.56.930
  7. Lee, D., Ono, W. T. and Esashi, M., 2000, "Cantilever with Integrated Resonator for Application of Scanning Probe Microscope," Sensors and Actuators A: Physical, Vol. 83, pp. 11-16. https://doi.org/10.1016/S0924-4247(99)00378-7
  8. Hazel, J. L. and Tsukruk, V. V., 1999, "Spring Constants of Composite Ceramic/gold Cantilevers for Scanning Probe Microscopy," Thin Solid Films Vol. 339, pp. 249-257. https://doi.org/10.1016/S0040-6090(98)00961-4
  9. Lee, M., McdDaniel, E. B. and J. W. P. Hsu, 1996, "An Impedance Based Non-contact Feedback Control System for Scanning Probe Microscopes," American Institute of Physics, Rev. Sci. Instrum. 67.
  10. Cozzens, R., 2002, CATIA Version 5: Release 8 & 9, Schroff Development Co.