Pressure fluctuations of the hot cathode ionization gauges at the stainless steel high vacuum chamber according to room temperature variations

스테인레스 스틸 고진공용기에 부착된 열음극 전리진공게이지의 주변 온도변화에 따른 압력요동

  • ;
  • ;
  • ;
  • ;
  • Ichiro Arakawa (Department of Physics, Gakushuin University)
  • 홍승수 (한국표준과학연구원 진공기술센터) ;
  • 임인태 (한국표준과학연구원 진공기술센) ;
  • 신용현 (한국표준과학연구원 진공기술센) ;
  • 정광화 (한국표준과학연구원 진공기술센) ;
  • Published : 2004.06.01

Abstract

Three hot cathode ionization gauges (HCIG) were used to study their responses of pressure measurements at the stainless steel high vacuum (HV) chamber around room temperatures. The pressure variations of HV measured by the three HCIGs between $20 ^{\circ}C$ and $26 ^{\circ}C$ were 5.0 %, 5.3 %, and 10.3 %, respectively. However, pressure differences between $21.5 ^{\circ}C$ and $22.5^{\circ}C$ were lower than the gauges uncertainty limits. The pressure variations of the HCIGs at the HV chamber was not directly affected by gauge characteristics, but pressure changes due to room temperatures.

스테인레스 스틸로 가공된 진공용기에 제조회사와 사양이 각각 다른 세 개의 열음극 전리진공게이지를 부착하고 실험실 온도변화에 따른 게이지들의 압력요동 특성을 조사하였다. 상온 근방의 최소 $20 ^{\circ}C$와 최대 $26 ^{\circ}C$에서의 압력 편차는 각각 IG1은 5.0 %, IG2는 5.3 %, 그리고 IG3는 10.3 %이었다. 그러나 온도 $21.5^{\circ}C$에서 $22.5^{\circ}C$구간에서의 압력변화는 모두 게이지들의 불확도 보다 작고 안정한 성능을 보였다. 본 연구결과 상온 고진공에서 열음극 전리진공게이지의 압력변화는 게이지 자체의 특성 때문이 아니고 실험실 온도 때문에 변하는 압력변화에 기인한 것임을 알 수 있었다.

Keywords

References

  1. Metrologia v.36 K.H.Chung;S.S.Hong;Y.H.Shin https://doi.org/10.1088/0026-1394/36/6/37
  2. 진공학 배석희;인상열;정광화;이영백;신용현
  3. J. Vac. Sci. Technol. v.A 12 Patrick J. Abbott;Patrick Looney
  4. Applied Surface Science v.169 K.Kokubun;M.Hirata https://doi.org/10.1016/S0169-4332(00)00787-X
  5. Vaccum v.49 K.Jousten