Hearing Test in Infant and Young Children Using MASTER

Multiple Auditory Steady-State Response(MASTER)을 이용한 유소아의 청력검사

  • 이학선 (연세대학교 의과대학 이비인후과학교실) ;
  • 김시찬 (순천향대학교 의과대학 이비인후과학교실)
  • Published : 2004.12.01

Abstract

This study was done to compare thresholds between those of Auditory Brainstem Response (ABR) with clicks or tonebursts and Multiple Auditory Steady-State Response (MASTER). The results would give a promising tool for evaluating frequency-specific hearing sensitivity in infants or young children. The correlation coefficient value between the click ABR thresholds and MASTER thresholds at carrier frequencies, 500, 1,000, 2,000 Hz, and 4,000 Hz was obtained at Pearson 0.91, 0.94, 0.93, and 0.91.

Keywords