Vacuum Assisted Auto-Lancing Technique for Capillary Blood Sampling on the Forearm with Minimized Pain

전완부위 최소통증 채혈을 위한 진공 자동 채혈기법

  • Park Mi Sook (Department of Biomedical Engineering, School of Medicine, Chungbuk National University) ;
  • Park Kyung Soon (Department of Biomedical Engineering, School of Medicine, Chungbuk National University) ;
  • Kim Kyung Ah (CKInternational Co. Ltd.) ;
  • Jun Myung Hee (Department of Nursing Science, Oriental Medical college, Taejeon University) ;
  • Kim Tae Im (Department of Nursing Science, Oriental Medical college, Taejeon University) ;
  • Lee Tae Soo (Department of Biomedical Engineering, School of Medicine, Chungbuk National University) ;
  • Cha Eun Jong (Department of Biomedical Engineering, School of Medicine, Chungbuk National University)
  • 박미숙 (충북대학교 의과대학 의공학교실) ;
  • 박경순 (충북대학교 의과대학 의공학교실) ;
  • 김경아 ((주)씨케이인터내셔널) ;
  • 전명희 (대전대학교 한의과대학 간호학과) ;
  • 김태임 (대전대학교 한의과대학 간호학과) ;
  • 이태수 (충북대학교 의과대학 의공학교실) ;
  • 차은종 (충북대학교 의과대학 의공학교실)
  • Published : 2004.12.01

Abstract

A new vacuum assisted auto-lancing technique is proposed to minimize pain. Specially designed lancing device was introduced, which applied -100mmHg right after skin puncture on the forearm. Sampled blood volumes were measured in 58 normal females. Mean volume of 464 samples was approximately 2.6$\muL$ and the frequency of more than 0.5$\muL$ was 86%. Thus the success rate of blood sugar test should also be the same when using modern glucose meters capable of testing with only 0.3~0.5$\muL$ of capillary blood. When pain scores were quantitatively evaluated by the visual pain measure, only 23% pain of the traditional finger sampling was experienced, demonstrating that capillary blood sampling was performed on the forearm with almost no pain. The present technique reduced pain to a great degree, though resampling might be unavoidable due to 14% of test failure rate estimated for modern glucose meters. However, minimized pain makes the present technique of great convenience for diabetic patients who need blood sampling a few times a day.

당뇨병 환자의 혈당검사시 통증을 최소화하는 새로운 진공 자동 채혈기법을 개발하였다. 특별히 고안, 제작된 채혈기구를 사용하여 전완부위에서 채혈침으로 피부를 관통한 직후 자동적으로 -100mmHg의 진공압력이 가해짐으로써 모세혈액이 흡입되는 자동화된 진공 채혈기법을 구현하였다. 진공 자동 채혈기법의 유용성을 검증하기 위하여 전완부위에서의 채혈용적을 측정하였다. 58명의 정상인을 대상으로 464회의 채혈을 시행한 결과 평균 채혈용적은 약 2.6$\muL$이었고 0.5$\muL$이상의 채혈빈도가 약 86%이었다. 최근 개발되고 있는 혈당계들은 0.3~0.5$\muL$의 모세혈액으로 혈당검사가 가능하므로 검사성공률 역시 약 86%로 추정되었다. 채혈통증 감소효과를 평가하기 위하여 손가락 및 전완부위 채혈 후 시각상사 통증척도를 사용하여 통증지수를 측정하였다. 전완부위 채혈시 통증은 손가락 채혈의 약 23%에 불과하였으며 거의 무통증 채혈이 가능하였다. 따라서 본 연구의 전완부위 진공 자동채혈기법은 전통적인 손가락 채혈에 비해 채혈통증이 25% 이하로 현저하게 감소하지만 검사실패율이 14%로써 경우에 따라 재채혈이 필요할 수 있다. 그러나 매일 수 차례 채혈해야 하는 만성 당뇨환자의 경우 검사기피를 방지할 수 있는 유용한 방법으로 생각된다.

Keywords

References

  1. J. D. Baum, 'Home monitoring of diabetic control', Arch Dis Children, Vol. 56, pp. 897-899, 1981
  2. J. H. Baumer, A. D. Edelsten, B. C. Howlett, C. Owens, C. A. Pennock, D. C. L. Savage, 'Impact of home blood glucose monitoring on childhood diabetes', Arch Dis Children, Vol. 57, pp. 195-199, 1982
  3. R. M. Cowett, L. D'Amico, 'Capillary(Heelstick) versus venous blood sampling for the determination of glucose concentration in the neonate', BioI Neonate, Vol. 62, pp 32-36, 1992
  4. D. P. Barker, N Rutter, 'Exposure to invasive procedures in neonatal intensive care unit admissions', Arch Dis Childhood, Vol. 72, pp. F48-F49, 1995
  5. N. Mclntosh, L. van Veen, H. Brameyer, 'Alleviation of the pain of heel prick in preterm infants', Arch Dis Childhood, Vol. 70, pp. FI77-F181, 1994
  6. M. E. Loveland, S. D. Carley, N. Cranfield, V. F. Hiller, K. Mackway-Jones, ' Assessment of the pain of blood sugar testing: a randomized controlled trial', Lancet, Vol. 354, pp. 921, 1999
  7. 김현식, 김경아, 전병희, 김태임, 정용현, 이태수, 차은종, '피부관통시 통증을 최소화하는 일회용 자동 모세혈액채취기구', 의공학회지, 제 23권, 제 4호, pp. 323-327, 2002
  8. D. D. Cunningham, T. P. Henning, E. B. Shain, D. F. Young, T. A. Elstrom, E. J. Taylor, S. M. Schroder, P. M. Gatcomb, and W. V. Tamborlane, 'Vacuum-assisted lancing of the forearm: an effective and less painful approach to blood glucose monitoring', Diabetes Technology & Therapeutics, Vol. 2, pp. 541-548, 2000
  9. A. J. Brenneman, D. G. Purcell, 'Vacuum assisted lancing device', U. S. Patent No. 6,152,942, 2000
  10. G. McGarraugh, D. Price, S. Schwartz, and R. Weinstein, 'Physiological influen ces on off-finger glucose testing', Diabetes Technology & Therapeutics, Vol. 3, pp. 367-376, 2001
  11. 차은종, 박미숙, 최강, 담학현, 차근식, 'Vacuum assisted auto-lancing device', PCT/KR2004/j000351, 2004
  12. 박미숙, 박경순, 김태임, 전병희, 이태수, 차은종, '당뇨환자 혈당검사를 위한 부위별 모세혈액 채혈량 비교', 제 28회 대한의용생체공학회 춘계학술대회, P-104, 2003
  13. 박미숙, 박경순, 김경아, 이재헌, 전명희, 김태임, 이태수, 차은종, '실험용 배지를 사용한 미량의 모세혈액 측정기법', 제 29회 대한의용생체공학회 추계학술대회, P-33, 2003
  14. Agency for Health Care Policy and Research, 'Acute pain management in adults: operative procedures', Quick reference guide for clinicians No. 1a, AHCPR Publication No. 92-0019, Public Health Service, U. S. Dept. of Health and Human Services, 1993
  15. T. M. Blumenfeld, G. K. Turi, W. A. Blanc, 'Recommended site and depth of newborn heel skin punctures based on anatomical measurements and histopathology', Lancet, Vol. 334, pp. 230-233, 1979
  16. G. McGarraugh, S. Schwartz, and R. Weinstein, 'Glucose measurements using blood extracted from the forearm and the finger', Techincal Report ART01022 Rev. c., TheraSense Inc., 2001