The Study on the Head and Neck Phantom for Quality Assurance of Intensity Modulated Radiotherapy

세기변조방사선치료의 정도관리를 위한 두경부 팬톰 제작에 관한 연구

  • Shin Dongho (Research Institute and Hospital, National Cancer Center) ;
  • Park Sung-Yong (Research Institute and Hospital, National Cancer Center) ;
  • Kim Joo Young (Research Institute and Hospital, National Cancer Center) ;
  • Lee Se Byeong (Research Institute and Hospital, National Cancer Center) ;
  • Cho Jung Keun (Research Institute and Hospital, National Cancer Center) ;
  • Kim Dae Yong (Research Institute and Hospital, National Cancer Center) ;
  • Cho Kwan Ho (Research Institute and Hospital, National Cancer Center)
  • 신동호 (국립암센터 양성자치료센터) ;
  • 박성용 (국립암센터 양성자치료센터) ;
  • 김주영 (국립암센터 양성자치료센터) ;
  • 이세병 (국립암센터 양성자치료센터) ;
  • 조정근 (국립암센터 양성자치료센터) ;
  • 김대용 (국립암센터 양성자치료센터) ;
  • 조관호 (국립암센터 양성자치료센터)
  • Published : 2005.03.01

Abstract

For the QA of IMRT treatment of head and neck cancer by using M3 (BrainLAB Inc. Germany), it is not easy to measure delivery dose exactly because the dose attenuation appears by the couch according to the position of table and gantry. In order to solve this problem, we fabricated head and neck phantom which would be implemented on the couch mount of Brain Lab Inc. We investigated dose attenuation by the couch and found the difference of dose distribution by the couch, in the applying this phantom to the clinic. After measurement, we found that point dose attenuation was 35% at maximum and dose difference was 5.4% for a point dose measurement of actual patient quality assurance plan.

m3 (BrainLAB Inc., Germany)를 이용한 두경부 IMRT의 정도관리에서 테이블과 갠트리의 위치에 따라 테이블에 의한 선량감쇠가 일어나 정확한 처방 선량을 측정할 수 없다. 이 문제를 해결하기 위해 두경부 종양의 세기변조방사선치료를 위해 Brain Lab사의 환자테이블 mount를 이용해 설치할 수 있는 원통형 두경부 팬톰을 제작하였다. 이를 이용하여 환자테이블에 의한 선량 감쇠를 측정하고 실제 임상에 적용함으로써 테이블에 의한 선량 감쇠로 인한 선량분포의 차이를 확인할 수 있었다. 측정결과 환자테이블에 의한 점 선량의 감쇠가 최대 약 35%가 났으며 실제 환자 치료계획에 대한 정도관리에서의 절대점 선량의 경우 5.4%의 선량차이를 나타냈다.