THREE DIMENSIONAL FINITE ELEMENT ANALYSIS OF INTERNALLY CONNECTED IMPLANT SYSTEMS

내부연결방식 임플랜트 시스템의 삼차원 유한요소법적 연구

  • Kim Yu-Lee (Department of Prosthodontics, College of Dentistry, Wonkwang University) ;
  • Cho Hye-Won (Department of Prosthodontics, College of Dentistry, Wonkwang University) ;
  • Lee Jai-Bong (Department of Prosthodontics, College of Dentistry, Seoul National University)
  • 김유리 (원광대학교 치과대학 보철학교실) ;
  • 조혜원 (원광대학교 치과대학 보철학교실) ;
  • 이재봉 (서울대학교 치과대학 보철학교실)
  • Published : 2006.02.01

Abstract

Statement of problem: Currently, there are some 20 different geometric variations in implant/abutment interface available. The geometry is important because it is one of the primary determinants of joint strength, joint stability, locational and rotational stability. Purpose: As the effects of the various implant-abutment connections and the prosthesis height variation on stress distribution are not yet examined this study is to focus on the different types of implant-abutment connection and the prosthesis height using three dimensional finite element analysis. Material and method. The models were constructed with ITI, 3i TG, Bicon, Frialit-2 fixtures and solid abutment, TG post, Bicon post, EstheticBase abutment respectively. And the super structures were constructed as mandibular second premolar shapes with 8.5 mm, 11 mm, 13.5 mm of crown height. In each model, 244 N of vertical load and 244 N of $30^{\circ}$ oblique load were placed on the central pit of an occlusal surface. von Mises stresses were recorded and compared in the crowns, abutments, fixtures. Results: 1. Under the oblique loading, von Mises stresses were larger in the crown, abutment, fixture compared to the vertical loading condition. 2. The stresses were increased proportionally to the crown height under oblique loading but showed little differences with three different crown heights under vertical loading. 3. In the crown, the highest stress areas were loading points under vertical loading, and the finish lines under oblique loading. 4. Under the oblique loading, the higher stresses were located in the fixture/abutment interface of the Bicon and Frialit-2 systems compared to the ITI and TG systems. Conclusions: The stress distribution patterns of each implant-abutment system had difference among them and adequate crown height/implant ratio was important to reduce the stresses around the implants.

Keywords

References

  1. Adell R. Lekholm U. Rockler B. Branemark PI. A 15-year study of osseointegrated im plants in the treatment of the edentulous jaw. J Oral Surg 1981: 10:387-416 https://doi.org/10.1016/S0300-9785(81)80077-4
  2. Lekholm U. van Steenberghe D. Herrmann I. Bolender C. Folmer T. Gunne J. et al. Osseointegrated implants in the treatment of partially edentulous jaws: A prospective 5-year multicenter study. Int J Oral Maxillofac Implants 1994;9:627-35
  3. Buser D. Mericske Stern R. Bernard JP. Behneke A, Behneke N. Hirt HP. et al. Long term evaluation of non submerged ITI implants. Part 1 :8-year life table analysis of a prospective multi-center study with 2359 implants. Clin Oral Implants Res 1997;8:161-72 https://doi.org/10.1034/j.1600-0501.1997.080302.x
  4. Laney WR, Jemt T. Harris D. Henry PJ. Kroger PHJ. Polizzi G. et al. Osseointegrated implants for single-tooth replacement: Progress report from a multicenter prospective study after 3 years. Int J Oral Maxillofac Implants 1994:9:49-54
  5. Levine RA, Clem DS III. Wilson TG Jr. Higginbottom F. Saunders SL. A multicenter retrospective analysis of the ITI implant system used for single-tooth replacements: Preliminary results at 6 or more months of loading. Int J Oral Maxillofac Implants 1997: 12:237-42
  6. Gomez-Roman G. Schulte W. d' Hoedt B. Axman-Krcmar D. The Frialit-2 implant system: Five-year clinical experience in single-tooth and immediately post-extraction applications. Int J Oral Maxillofac Implants 1997: 12: 299-309
  7. Jemt T. Laney WR, Harris D. Henry PJ. Krogh PHJ Jr. Polizzi G, et al. Osseointegrated implants for single tooth replacement: A Iryear report from a multicenter prospective study. Int J Oral Maxillofac Implants 1991 :6:29-36
  8. Albrektsson T. Zarb G. Worthington P. Eriksson AR. The long-term efficacy of currently used dental implants: a review and proposed criteria of success. Int J Oral Maxillofac Implants 1986: 1: 11-25
  9. Sones AD. Complications with osseointegrated implants. J Prosthet Dent 1989: 62:581-5 https://doi.org/10.1016/0022-3913(89)90084-X
  10. Zarb GA, Schmitt A. The longitudinal clincal effectiveness of osseointegrated implants: The Tronto study, Part III. Problems and complications encountered. J Prosthet Dent 1990:64:185-90 https://doi.org/10.1016/0022-3913(90)90177-E
  11. Jemt T, Linden B, .Lekholm U. Failures and complications in 127 consecutively placed fixed partial prostheses supported by Branemark implants: From prosthetic treatment to first annual check up. Int J Oral Maxillofac Implants 1992: 7: 40-4
  12. Jemt T, Lekholm U. Oral implant treatment in posterior partially edentulous jaws: A 5 year follow up study. Int J Oral Maxillofac Implants 1993:8:635-40
  13. Kallus T, Bessing C. Loose gold screws frequently occur in full arch fixed prostheses supported by osseointegrated implants after 5 years. Int J Oral Maxillofac Implants 1994:9169-78
  14. Jemt T. Fixed implant supported prostheses in the edentulous maxilla: A five-year follow-up report. Clin Oral Implants Res 1994:5:142-7 https://doi.org/10.1034/j.1600-0501.1994.050304.x
  15. Wie H. Registration of localization, occlusion and occluding materials for failing screw-joints in the Branemark implant system. Clin Oral Implants Res 1995:6:47-53 https://doi.org/10.1034/j.1600-0501.1995.060106.x
  16. Becker W, Becker BE. Replacement of maxillary and mandibular molars with single endosseous implant restorations: A retrospective study. J Prosthet Dent 1995: 74:51-5 https://doi.org/10.1016/S0022-3913(05)80229-X
  17. Balshi TJ. First molar replacement with osseointegrated implants Quintessence Int 1990;21:61-5
  18. Balshi TJ. Hernandez RE. Pryszlak C. Rangert B. A comparative study of one implant versus two replacing a single molar. Int J Oral Maxillofac Implants 1996; 11: 372-8
  19. Binon pp. Implants and components: Entering the new millennium. Int J Oral Maxillofac Implants 2000; 15: 76-94
  20. Norton MR. An invitro evaluation of the strength of an internal conical interface compared to a butt joint interface in implant design. Clin Oral Implants Res 1997;8:290-8 https://doi.org/10.1034/j.1600-0501.1997.080407.x
  21. Norton MR. Assessment of cold welding properties of the internal conical interface two commercially available implant systems. J Prosthet Dent 1999;81:159-66 https://doi.org/10.1016/S0022-3913(99)70243-X
  22. Felton DA. Cemented versus screw-retained implant prostheses: Which is better- Int J Oral Maxillofac Implants 1999; 14: 138-9
  23. Levine RA. Clem DS. Wilson TGJr. Higginbottom F. Saunders SL. A multicenter retrospective analysis of the ITI implant system used for single-tooth replacements: Results of loading for 2 of more years. Int J Oral Maxillofac Implants 1999: 14:516-20
  24. Penny RE. Kraal JH. Crown to root ratio: its significance in restorative dentistry. J Prosthet Dent 1979;42:34-8 https://doi.org/10.1016/0022-3913(79)90327-5
  25. Turesky JD. Shepherd NJ. Morgan VJ. Muftu A. A simple prosthetic approach using cement-retained implant prosthesis after surgical treatment of ameloblastoma. Implant Dent 1999;8:407-12 https://doi.org/10.1097/00008505-199904000-00015
  26. Weinstein AM. Klawitter JJ. Anand SC. Schuessler R. Stress analysis of porous rooted dental implants. J Dent Res 1976: 55:772-7 https://doi.org/10.1177/00220345760550051001
  27. Major M Sch Jr. Wheeler's dental anatomy, physiology and occlusion: 7th ed WB Saunders Co 1993:218-31
  28. Borchers L, Reichart P. Three-dimensional stress distribution around a dental implant at different stages of interface development. J Dent Res 1983:62:155-9 https://doi.org/10.1177/00220345830620021401
  29. Collings EW. The physical metallurgy of titanium alloys. Metals park(OH) : Americal society for metals: 1984
  30. Craig RG. Restorative dental materials. 8th ed. St. Louis (MO) :Mosby: 1989. P.84
  31. Dario LJ. How occlusal forces change in implant patients. A clinical research report. J Am Dent Associ 1995:126:1130
  32. Misch CE, Badez MW. Implant protected occlusion: A Biomechnical rationale. Compendium 1994: 15: 1330
  33. Le Gall MG, Jauret JF, Saadoun AP. Mastication forces and implant-bearing surface. Pract Periodontics Aesthet Dent 1994:6:37
  34. Weinberg LA. Reduction of implant loading with therapeutic biomechanics. J Implant Dent 1998:7:277 https://doi.org/10.1097/00008505-199807040-00005
  35. Richter EJ. In vivo vertical forces on implants. Int J Oral Maxillofac Implants 1995:10:99
  36. Misch CE, Bidez MW. Implant protected occlusion. Pract Periodontics Aesthet Dent 1995:7:25
  37. McGlumphy EA, Mendel DA, Holloway JA. Implant screw mechanics. Dent Clin North Am 1998:42:71-89
  38. Jemt T, Pettersson P. A 3-year follow-up study on single implant treatment. J Dent 1993:21:203-8 https://doi.org/10.1016/0300-5712(93)90127-C
  39. Jansen VK, Conrads G, Richter EJ. Microbial leakage and marginal fit of the implant-abutment interface. Int J Oral Maxillofac Implants 1997; 12:527-40
  40. McGlumphy EA. Robinson DM. Mendel DA. Implant superstructures: a comparison of ultimate failure force. Int J Oral Maxillofac Implants 1992; 7: 35-9
  41. Balfour A. O' Brien GR. Comparative study of antirotational single tooth abutments. J Prosthet Dent 1995; 73: 36-43 https://doi.org/10.1016/S0022-3913(05)80270-7
  42. Sutter F, Weber HP, Sorensen J. Belser U. The new restorative concept of the ITI dental implant system: design and engineering. Int J Periodont Rest Dent 1993;13:409-31
  43. Nevins M. The placement of mandibular posterior implants. In: Nevins M(eds). Implant Therapy. Clinical Approaches and Evidence of Success Chicago: Quintessence, 1998: 145-56
  44. Lekholm U, Gunne J, Henry P. Higuchi K Linden U, Bergstrom C, van Steenberghe D. Survival of the Branemark implant in partially edentulous jaws: A 10-year prospective multicenter study. Int J Oral Maxillofac Implants 1999; 14: 639-45
  45. Stellingsma C. Meijer HJ. Raghoebar GM. Use of short endosseous implants and an overdenture in the extremely resorbed mandible: a five year retrospective study. J Oral Maxillofac Surg 2000: 58: 3827:discussion 387-8 https://doi.org/10.1016/S0278-2391(00)90918-2
  46. Bidez MW. Misch CE: Force transfer in implant dentistry: basic concepts and principles. Oral Implantol 1992:18:264-74
  47. Phillips RW. Skinner' s science of dental materials. 9th ed. Philadelphia: WB Saunders. 1991: 15
  48. Ablowitz S. Burke JJ. Hiltz RH. Titanium in Industry. New York: Van Nostrand Co Inc. 1955
  49. Steinemann SG. Mausli PA. SzmuklerMoncler S. Semlitzsch M. Pohler O, Hintermann HE. Perren SM. Betatitanium alloy for surgical implants. In: Froes FH. Caplan I(eds) .Titanium '92.Science and Technology. The Minerals, Metals& Materials Society. 1993:2689-96
  50. Sakaguchi RL. Borgersen SE. Nonlinear contact analysis of preload in dental implant screws. Int J Oral Maxillofac Implants 1995: 10: 295-302