FATIGUE LIFE ESTIMATION OF IMPLANT USING A FINITE ELEMENT METHOD

유한요소법을 이용한 치아 임플랜트 피로수명 예측

  • Published : 2006.08.01

Abstract

Purpose : The purpose of this study is to use finite element analysis to predict the fatigue life of an implant system subjected to fatigue load by mastication (chewing force). The reliability and the stability of implant system can be defined in terms of the fatigue strength. Not only an implant is expensive but also it is almost impossible to correct after it is inserted. From a bio-engineering standpoint, the fatigue strength of the dental implant system must be evaluated by simulation (FEA). Material and Methods Finite element analysis and fatigue test are performed to estimate the fatigue strength of the implant system. Mesh of implant is generated with the actual shape and size. In this paper, the fatigue strength of implant system is estimated. U-fit (T. Strong, Korea, internal type). The stress field in implant is calculated by elastic-plastic finite element analysis. The equivalent fatigue stress, considering the contact and preload stretching of a screw by torque for tightening an abutment, is obtained by means of Sine's method. To evaluate the reliability of the calculated fatigue strength, fatigue test is performed. Results: A comparison of the calculated fatigue strength with experimental data showed the validity and accuracy of the proposed method. The initiation points of the fatigue failure in the implant system exist in the region of high equivalent fatigue stress values. Conclusion: The above proposed method for fatigue life estimation tan be applied to other configurations of the differently designed and improved implant. In order to prove reliability of prototype implant, fatigue test should be executed. The proposed method is economical for the prediction of fatigue life because fatigue testing, which is time consuming and precision-dependent, is not required.

Keywords

References

  1. Ibrahim A. Influence of occlusal forces on stress distribution in preloaded dental implant screw. J Prosthet Pent 2004; 91: 319-325 https://doi.org/10.1016/j.prosdent.2004.01.016
  2. Bulent E. Numerical analysis of a dental implant system in three-dimension. Advances in Engineering Software 2002; 33:109-113 https://doi.org/10.1016/S0965-9978(01)00056-4
  3. Fatigue test for endosseous dental implants. ISO14801 The International Organization for Standardization 2003
  4. Funch HO. Metal fatigue in engineering. 1980;173-186
  5. Matthew JD. Titanium a technical guide: ASM interanational; 2000; 105
  6. Kazuaki S. Titanium and Titanium alloys. In: Databook on fatigue strength of metallic materials. Amsterdam: Elsevier;1996; 2:1746-64