Abstract

Previous trials for development of a pulsatile-Extracorporeal Life Support had some defects such as insufficient blood flow, high pressure at its membrane oxygenator and the high risk of blood cell damage. To solve those problems of previous pulsatile-ECLSs, we suggest dual pulsatile blood pump structure for the new pulsatile-ECLS. Two pulsatile pumps areconnected in a parallel manner and this new structure raises the inflow capacity and efficiency and it decreases the high blood pressure at membrane oxygenator. In in-vitro experiments, The Energy Equivalent Pressure Increment(EEP inc.) was 10%, and it showed that its pulsatilty was $5{\sim}10$ times higher than other commercial ECLS In in-vivo experiments, we had applied a new pulsatile-ECLS to 30 Kg pigs and a new pulsatile-ECLS couldsupport high blood flow and pulsatility above 2 L/min, 10% EEP inc.

Keywords

References

  1. J.K. Boucher, L.W. Rudy, and L.H. Edmunds, 'Organ blood flow during pulsatile cardio pulmonary bypass,' J. Appl. Physiol., vol.36, no.1, pp.86-90, 1974 https://doi.org/10.1152/jappl.1974.36.1.86
  2. A.H. Harken, 'The influence of pulsatile perfusion on oxygen uptake by the isolated canine hind limb,' J. Thorac. Cardiovasc Surg., vol.70, no.2, pp.237-241,1975
  3. B.Y. Chiang, C.X. Ye, X.D. Gau, Y. Zhou, J.P. Hong, and Y.S. Wang, 'Effect of pulsatile reperfusion on myocardial high energy phosphates following global ischemia,' ASAIO Trans., vol.37, pp.M485-M487, 1991
  4. P.R. Hichey, M.J. Buckley, and D.M. Philbin, 'Pulsatile and nonpulsatile caridopulmonary bypass review of a counterproductive controversy,' Ann. Thorac. Surg., vol.36, no.6, pp.720-737, 1983 https://doi.org/10.1016/S0003-4975(10)60286-X
  5. M.P. Anstadt, M. Tedder, S.S. Hegde, et al. 'Pulsatile versus nonpulsatile reperfusion improves cerebral blood flow after cardiac Arrest,' Ann Thorac. Surg., vol. 56, pp.453-461, 1993 https://doi.org/10.1016/0003-4975(93)90879-M
  6. R.B. Shepard, and J.W. Kirklin, 'Relation of pulsatile flow to oxygen consumption and other variables during cardiopulmonary bypass,' J. Thorac. Cardiovasc Surg., vol.58, no.5, pp.694-702, 1969
  7. J.M. Sanderson, G. Wright, and F.W. Sims, 'Brain damage in dogs immediately following pulsatile and non-pulsatile blood flows in extracorporeal circulation,' Thorax, vol.27, pp.275-286, 1972 https://doi.org/10.1136/thx.27.3.275
  8. J. Dunn, M.M. Kirsh, J. Harness, M. Carroll, J. Straker, and H. Sloan, 'Hemodynamic, metabolic, and hematologic effects of pulsatile cardiopulmonary bypass,' J. Thorac. Cardiovasc. Surg., vol.68, no.1, pp.138-147, 1974
  9. E. Dapper, H. Neppl, G. Woznicak, et al. 'Effects of pulsatile and nonpulsatile perfusion mode during extracorporeal circulation - a comparative clinical study,' J.Thorac. Cardiovasc. Surg, vol.40, pp.345-351, 1992 https://doi.org/10.1055/s-2007-1020178
  10. K. Minami, K. Vyska, and R. Korfer, 'Role of the carotid sinus in response of integrated venous system to pulsatile and nonpulsatile perfusion,' J. Thorac. Cardiovasc. Surg, vol.104, no.6, pp.1639-1646, 1992
  11. K. Nakayama, T. Tamiya, K. Yamamoto, et al. 'High-amplitude pulsatile pump in extracorporeal circulation with particular reference to hemodynamics,' Surgery, vol.54, pp.798-809, 1963
  12. J.K. Trinkle, N.E. Helton, R.E. Wood, and L.R. Bryant, 'Metabolic comparison of a new pulsatile pump and a roller pump for cardiopulmonary bypass,' J. Thorac. Cardiovasc. Surg., vol.58, no.4, pp.562-569, 1969
  13. T. Eiki, N. Yoshinari, T. Yoshiyuki, et al. 'Hemolysis test of a centrifugal pump in a pulsatile mode: The effect of pulse rate an RPM variance,' Artif Organs, vol.21, no.12, pp.1284-1287, 1997 https://doi.org/10.1111/j.1525-1594.1997.tb00489.x
  14. A. Undar, T. Masai, O.H. Frazier, and C.D. Frazier, 'Pulsatile and nonpulsatile flows can be quantified in terms of energy equivalent pressure during cardiopulmonary bypass for direct comparisons,' ASAIO J., vol.45, pp.610-614, 1999 https://doi.org/10.1097/00002480-199911000-00017
  15. A. Undar, O.H. Frazier, and C.D. Frazier, 'Defining pulsatile perfusion: quantification in terms of energy equivalent pressure,' Artif Organs, vol.23, no.8, pp.712-716, 1999 https://doi.org/10.1046/j.1525-1594.1999.06409.x
  16. R.H. Paul, J.B. Mortimer, and M.P. Daniel, 'Pulsatile and nonpulsatile cardiopulmonary bypass: review of a counterproducitve controversy,' Ann. Thorac. Surg., vol.36, no.6, pp.720-737, 1983 https://doi.org/10.1016/S0003-4975(10)60286-X
  17. A. Undar, T. Masai, S.Q. Yang, et al. 'Pulsatile perfusion improves regional myocardial blood flow during and after hypothermic cardiopulmonary bypass in a neonatal piglet model,' ASAIO J., vol.48, pp.90-95, 2002 https://doi.org/10.1097/00002480-200201000-00017
  18. A. Undar, T. Masai, E.A. Beyer, J. Goddard-Finegold, M.C. McGarry, and C.D. Fraser, 'Pediatric physiologic pulsatile pump enhances cerebral and renal blood flow during and after cardiopulmonary bypass,' Artif. Organs., vol.26, no.11, pp.919-923, 2002 https://doi.org/10.1046/j.1525-1594.2002.07127.x
  19. Y. Nose, K. Kawahito, and T. Nakazawa, 'Can we develop a nonpulsatile permanet rotary blood pump? yes we can,' Artif. Organs., vol.20, no.6, pp.467-474, 1996 https://doi.org/10.1111/j.1525-1594.1996.tb04466.x
  20. T.P. Green, P. kreismer, R.H. Steinhorn, N.R. Payne, R.J. Irmiter, and C.L. Meyer, 'Comparison of pressure-volume-flow relationships in centrifugal and roller pump extracorporeal membrane oxygenation systems for neonates,' ASAIO Trans., vol.37, no.4, pp.572-576, 1991
  21. J.B. Atkinson, P. Emerson, R. Wheaton, and C.M. Bowman, 'A simplified method for autoregulation of blood flow in the extracorporeal membrane oxygenation circuit,' J. Pediatr. Surg., vol.24, no.3, pp.251-252, 1989 https://doi.org/10.1016/S0022-3468(89)80005-3
  22. K. Setz, K. Kesser, R.J. Kopotic, and J.D. Cornish, 'Comparison of a new venous control device with a bladder box system for use in ECMO,' ASAIO J., vol.38, no.4, pp.835-840, 1992 https://doi.org/10.1097/00002480-199238040-00018
  23. R.H. Steinhorn, B. Isham-Schopf, C. Smith, and T.P. Green, 'Hemolysis during ling-term extracorporeal membrane oxygenation,' J. Pediatr, vol.115, no.4, pp.625-30, 1989 https://doi.org/10.1016/S0022-3476(89)80299-9
  24. H. Schima, L. Huber, D. Melvin, et al. 'Effect of stationary guiding vanes on improvement of the washout behind the rotor in centrifugal blood pumps,' ASAIO J., vol.38, no.3, pp.M220, 1992 https://doi.org/10.1097/00002480-199207000-00024
  25. H.S. Lee, Y.R. Rho, H.S. Lee, et al. 'In-vivo evaluation of the pulsatile ECLS system,' J. Artif . Organ., vol.6, pp25-29, 2003 https://doi.org/10.1007/s100470300004
  26. Y.R. Rho, H. Choi, J.C. Lee, et al. 'Applications of the pulsatile flow versatile ECLS: In vivo Studies,' Int. J. Artif. Organs., vol.26, pp.428-435, 2003 https://doi.org/10.1177/039139880302600509
  27. J. Chung, W.E. Kim, J.J. Lee, et al. 'Assessment and improvement of the system efficiency for the moving-actuator type biventrifucular assist devices,' Artif. Organs., vol.28, no.6, pp.549-555, 2004 https://doi.org/10.1111/j.1525-1594.2004.07266.x
  28. R.B. Shepard, D.C. Simpson, and J.F. Sharp, 'Energy equivalent pressure,' Arch. Surg., vol.93, pp.730-740, 1966 https://doi.org/10.1001/archsurg.1966.01330050034005
  29. H.S. Son, K. Sun, Y.H. Fang, et al. 'The effect of pulsatile versus non-pulsatile extracorporeal circulation on the pattern of coronary artery blood flow during cardiac arrest,' Int. J. Artif. Organs., vol.28, no.6, pp.609-616, 2005 https://doi.org/10.1177/039139880502800610
  30. T.S. Kim, K. Sun, K.B. Lee, et al. 'Application of a pressure-relieving air compliance chamber in a single-pulsatile extracorporeal life support system: an experimental study,' Artif. Organs., vol.28, no.12, pp.1106-1109, 2004 https://doi.org/10.1111/j.1525-1594.2004.07341.x
  31. H.M. Kim, J.S. Shin, K. Sun, et al. 'Evaluation of the pulsatile ECLS system in heart failure model An animal experiment,' Abstract book of the 9the Annual Meeting of the Asian Society for Cardiovascular Surgery, pp.384, 2001
  32. J. Sands, D. Glidden, W. Jacavage, and B. Jones, 'Difference between delivered and prescribed blood flow in hemodialysis,' ASAIO J., vol.42, pp.74, 1996
  33. C.T. Ducko, M.L. McGregor, G. Rosenberg, and W.S. Pierce, 'The effect of valve type and drive line dP/dt on hemolysis in the pneumatic ventricular assist device,' Artif. Organs., vol.18, no.6, pp.454-460, 1994 https://doi.org/10.1111/j.1525-1594.1994.tb02232.x
  34. S.W. Choi, K.W. Nam, J. Chung, J.C. Lee, C.M. Hwang, Y.S. Won, K. Sun, B.G. Min, 'Development of a new pulsatile extracorporeal life support device incorporating a dual pulsatile blood pump,' Int. J. Artif. Organs., vol.29, no.6, pp.583-590, 2006 https://doi.org/10.1177/039139880602900607