Characteristics of Indium-Tin-Oxide Electrode for Continuous-flow PCR Chip

연속흐름 중합효소연쇄반응칩 제작을 위한 인듐 산화막 전극의 특성분석

  • Published : 2007.03.01

Abstract

We propose glass and PDMS (polydimethylsiloxane) chips for DNA amplification with continuous-flow PCR (polymerase chain reaction). The PDMS microchannel was fabricated using a negative molding method for sample injection. Three heaters and sensors of ITO (indium-tin-oxide) thin films were fabricated on glass chip. ITO heaters and sensors were calibrated accurately for the temperature control of the liquid flow. ITO heater generated stable heat versus applied power. ITO sensor resistance was changed linearly versus temperature increase as a RTD (resistance temperature detector) sensor. As a result, we enable precision temperature control of continuous-flow PCR chip. Using the continuous-flow PCR chip DNA plasmid pKS-GFP 720 bp was successfully amplified.

Keywords

References

  1. Ivonne SchneegaB, Johann Michael Kohler, 'Flow-through polymerase chain reactions in chip thermocycler', Molecular Biotechnology, vol. 82, pp. 101-121, 2001 https://doi.org/10.1016/S1389-0352(01)00033-2
  2. Q. Xiang, B. Xu, R. Fu, and D. Li, 'Real Time PCR on Disposable PDMS Chip with a Miniaturized Thermal Cycler', Biomedical Microdevices, vol. 7:4, pp. 273-279, 2005 https://doi.org/10.1007/s10544-005-6069-8
  3. Martin U. Kopp, Andrew J. de Mello, Andreas Manz, 'Chemical amplification : continuous-flow PCR on a chip', Science vol. 280, pp. 1046-1047, 1998 https://doi.org/10.1126/science.280.5366.1046
  4. Ivonne sehneegaB, Reiner Brautigam and Johann Michael Kohler, 'Miniaturized flow-through PCR with different template types in a silicon chip thermocycler', Lab on a chip vol. 1, pp. 42-49, 2001 https://doi.org/10.1039/b103846j
  5. Jeong Ah Kim, Ji Youn Lee, Shimyoung Seong, Seung Hwan Cha, Seung Hwan Lee, Jae Jeong Kim, Tai Hyun Park, 'Fabrication and characterization of a PDMS-glass hybrid continuous-flow PCR chip, Biochemical Engineering Journal vol. 29, pp. 91-97, 2006 https://doi.org/10.1016/j.bej.2005.02.032
  6. T. Fukuba, T. Naganuma and T. Fujii, 'Microfabricated flow-through PCR device for in situ gene analysis in extreme environments, MicroTAS, pp. 725-72B, 2003
  7. Tatsuhiro Fukuba, Takatoki Yamamoto, Takashi Naganuma, Teruo Fujii, 'Microfabricated flow-through device for DNA amplification-towards in situ gene analysis' Chemical Engineering Journal vol. 101, pp. 151-156, 2004 https://doi.org/10.1016/j.cej.2003.11.016
  8. 최용락, 정수열, 이영춘 역, 'PCR 실험의 원리와 응용', 세종출판사, 2003