Determination of Output Factors for the Gamma Knife using a Radiophotoluminescent Glass Rod Detector

유리선량계를 이용한 감마나이프의 출력인자 결정

  • Rah, Jeong-Eun (Department of Biomedical Engineering, College of Medicine, The Catholic University of Korea) ;
  • Suh, Won-Seop (Gamma Knife Center, Kyung Hee University Medical Center) ;
  • Shin, Dong-Oh (Department of Radiation Oncology, College of Medicine, Kyung Hee University) ;
  • Kim, Hee-Sun (Radiation Health Research Institute) ;
  • Suh, Tea-Suk (Department of Biomedical Engineering, College of Medicine, The Catholic University of Korea)
  • 라정은 (가톨릭대학교 의과대학 의공학교실) ;
  • 서원섭 (경희의료원 감마나이프실) ;
  • 신동오 (경희대학교 의과대학 방사선종양학교실) ;
  • 김희선 (한국수력원자력㈜ 방사선보건연구원) ;
  • 서태석 (가톨릭대학교 의과대학 의공학교실)
  • Published : 2007.03.31

Abstract

The purpose of study Is to Investigate whether glass rod detector (GRD) would be suitable for dosimeter of radiotherapy units. A GRD Is used for the measurement of the ou4put factors and x-axis beam profiles from Gamma Knife. The output factors measured with GRD from the 14, 8 and 4mm collimators relative to the 18mm collimator are $0.980{\pm}0.013,\;0.949{\pm}0.013\;and\;0.872{\pm}0.012$, respectively. The output factors obtained with a GRD are within 1.0% In good agreement with the values recommended by the manufacture. The full width at half maximum (FWHM) of x-axis beam profiles measured with GRD are 5.9mm at a 4mm collimator.

본 연구에서는 유리선량계가 방사선 치료장비의 선량계로서의 적정성 여부를 확인하고자 감마나이프의 출력인자 및 X축 측방선량분포를 측정하였다. 출력인자는 가장 큰 콜리메이터인 18mm의 측정값을 기준으로 하여 14, 8 그리고 4mm에 대해 상대적으로 측정하였고 그 결과 각각 $0.980{\pm}0.013,\;0.949{\pm}0.013$ 그리고 $0.872{\pm}0.012$로 나타났다. 이는 제작회사에서 제공하고 있는 표준데이터와 모든 콜리메이터에서 1.0% 이내에서 잘 일치하는 결과이다. 또한 유리선량계로 측정한 4mm 콜리메이터의 x축 측방선량분포의 반치폭은 5.9mm였다.